Then examine the parameters in Acqpars 'ased' mode (click 'pulse' icon), or type 'ased';
Most of the default parameters are suitable, however it's useful to compare values in the fields against those proposed by the original parameter file and the pulseprogram.
CNST4 - 1JNH value (≥93 Hz, used to calculate INEPT transfer delays)
CNST12 - multiplicity factor (NH: 4; NH/NH2: 8)
D1: recyle delay (~1 s for protonated samples, ~2-3 s for perdeuterated samples)
DS: 32-128 'dummy' scans that are not recorded ; allows system to reach steady state equilibration.
NS: multiple of 4; increase for increased signal to noise ( S/N increases as √NS )
VCLIST: Loop counter list for T2 delays. The length of each CMPG loop is given by D31 (typically ~8 ms). To minimize sample heating, set maximum loop counter such that the corresponding T2 delay is ≤250 ms;
-DLABEL_CN : enable 13C decoupling during 15N evolution for 13C,15N-labeled samples (default)
P1 - 1H 90º high power pulse (calibrated with calibo1p1 or BioTop)
SPdB1/SPW1 - power level [dB] for 1H H2O flip-back shaped pulse (calibrated with BioTop)
CNST21 - 13CO offset (~173 ppm, calibrated with BioTop) for 13C decoupling
CNST22 - 13CA offset (~53 ppm, calibrated with BioTop) for 13C decoupling
P21 - 15N 90º high power pulse (calibrated with BioTop) at PLW3, high power
P30 - 15N 180º pulse for CPMG spin-lock (normally ~6.4*P21) at PLW23, reduced power
PLW23 - Power for 15N 180º pulse for CPMG spin-lock, ~PLW3+10dB