Then examine the parameters in Acqpars 'ased' mode (click 'pulse' icon), or type 'ased';
Most of the default parameters are suitable, however it's useful to compare values in the fields against those proposed by the original parameter file and the pulseprogram.
CNST4 - 1JNH value (≥93 Hz, used to calculate INEPT transfer delays)
CNST12 - multiplicity factor (NH: 4; NH/NH2: 8)
D1: recyle delay (~1 s for protonated samples, ~2-3 s for perdeuterated samples)
DS: 32-128 'dummy' scans that are not recorded ; allows system to reach steady state equilibration.
NS: multiple of 4; increase for increased signal to noise ( S/N increases as √NS )
VDLIST: T1 delay list (use multiples of 10 ms). Click on 'E' to edit. The longest delay should be ~2-3s; it determines the temperature compensation.
-DLABEL_CN : enable 13C decoupling during 15N evolution for 13C,15N-labeled samples (default)
P1 - 1H 90º high power pulse (calibrated with calibo1p1 or BioTop)
SPdB1/SPW1 - power level [dB] for 1H H2O flip-back shaped pulse (calibrated with BioTop)
CNST21 - 13CO offset (~173 ppm, calibrated with BioTop) for 13C decoupling
CNST22 - 13CA offset (~53 ppm, calibrated with BioTop) for 13C decoupling
P21 - 15N 90º high power pulse (calibrated with BioTop)