If required, turn on the computer for the Agilent UV-VIS Cary-60 instrument and then turn on the instrument and wait until the LED stops blinking.
Open the Agilent software and wait until the instrument fully initializes.
To select experimental parameters, select Setup in the Agilent software.
Select wavelength range if desired (default is 190-1100 nm)
Click the baseline tab, select “Baseline correction” and select OK.
Click on “Zero” for lamp intensity correction. When performing this procedure, the sample compartment has to be empty.
Click on “Baseline”.
Fill an empty cuvette with DI water (blank) and another cuvette with the silver nanoparticle (AgNP) sample
Wipe the blank cuvette with a Kimwipe and inspect the surface for any debris
Place the blank cuvette in the sample compartment.
Close the lid of the chamber
Click OK to scan the baseline
Remove the blank
Wipe the sample cuvette with a Kimwipe
Insert the sample cuvette into the holder
Click Start to run the scan
When finished with measurement, select Finish
If processing multiple samples with the same parameters, select OK after each run.
If processing a large batch of samples (>20), scan a blank cuvette every 7 samples as a quality control (QC) procedure
After the scan is complete, save the file under a unique name (see data management section)
After finishing the experiments, close the software.
Leave both the instrument and the computer on.
Representative UV-VIS absorbance for the AgNP solution using an excitation wavelength of 190nm is shown in Fig 3
A clear peak at 395nm is observed for the AgNP solution prepared using this protocol (no custom emission filters or signal processing was performed).