If acquisition of a 1D spectrum requires substantial signal averaging, this SOP may not be appropriate.
The addition of echo detection increases the minimum number of scans required per time increment to 3. If the sensitivity of the sample is sufficient to obtain adequate signal with one scan and there is no significant background signal, instead follow the SOP for measuring T1 via saturation recovery without echo detection.
If acquisition of a 1D spectrum requires substantial signal averaging, this SOP may not be appropriate.
This SOP is intended for a beginning user who has been trained on basic spectrometer safety and principles.
This SOP is written for acquisition using a Bruker console and TopSpin 3.6.2. It may be adapted to other TopSpin versions where appropriate. The basic process is the same on TopSpin 4.x, but requires a NEO-compatible pulse sequence. The sequence satrect1_echo.neo.nmrfam is compatible.