Place samples in the Bullet Blender homogenizer. Homogenize for 30s to 1min on setting 8. Repeat this step as many times as needed until the tissue is fully homogenized. Incubate samples on ice between each homogenization cycle.
The number of homogenization cycles will depend the size of the tissue and the tissue type. If possible, homogenize some test samples first to identify the most appropriate homogenization procedure.
Leave at least three empty spaces between tubes in the homogenizing. Fully loading the bullet blender will reduce the effectiveness of the homogenizer and this step will take much longer.
Using the homogenizer on full speed (9 or 10) will risk having the sample tubes pop oven and spilling TRIzol‱. This not only creates a safety hazard but also results in the loss of precious samples.
Homogenized samples can be stored at -80C for future extraction.