May 27, 2022

Public workspaceQuantification of the SARS-CoV-2 using Nanotrap Particles®, the QIAcuity™ Digital PCR System, and GT-Digital SARS-CoV-2 Wastewater Surveillance Assay

  • 1University of Idaho
Icon indicating open access to content
QR code linking to this content
Protocol CitationSolana Narum, Thibault Stalder, Erik Coats, Eva Top 2022. Quantification of the SARS-CoV-2 using Nanotrap Particles®, the QIAcuity™ Digital PCR System, and GT-Digital SARS-CoV-2 Wastewater Surveillance Assay. protocols.io https://dx.doi.org/10.17504/protocols.io.e6nvwk8b2vmk/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it’s working.
Created: May 02, 2022
Last Modified: May 27, 2022
Protocol Integer ID: 61838
Keywords: SARS-CoV-2, Wastewater, Wastewater Based Epidemiology
Disclaimer
FOR INFORMATIONAL PURPOSES ONLY; USE AT YOUR OWN RISK

The protocol content here is for informational purposes only and does not constitute legal, medical, clinical, or safety advice, or otherwise; content added to protocols.io is not peer-reviewed and may not have undergone a formal approval of any kind. Information presented in this protocol should not substitute for independent professional judgment, advice, diagnosis, or treatment. Any action you take or refrain from taking using or relying upon the information presented here is strictly at your own risk. You agree that neither the Company nor any of the authors, contributors, administrators, or anyone else associated with protocols.io, can be held responsible for your use of the information contained in or linked to this protocol or any of our Sites/Apps and Services.
Abstract
This protocol was developed in an effort to serve as a timely and efficient method for the surveillance of the SARS-CoV-2 in primary influent wastewater samples. This process describes viral concentration via Nanotrap particles, RNA extraction using the Qiagen AllPrep PowerViral DNA/RNA kits with the Qiacube Connect, and quantification of the N1 and N2 genes in SARS-CoV-2 using the GT Digital SARS-CoV-2 Wastewater Surveillance for QIAcuity.

To compile the entire process from beginning to the end some sections were taken from the AllPrep PowerViral DNA/RNA Kit (Qiagen) and the GT Digital SARSCoV- 2 Wastewater Surveillance for QIAcuity v1.0 (GT Molecular) handbooks as well as the Ceres Nano Manual Nanotrap Wastewater Protocol using Qiagen AllPrep PowerViral DNA/RNA Kit protocol (Ceres Nano).
Materials
Equipment:

Equipment
DynaMag-2
NAME
Magnet
TYPE
Invitrogen
BRAND
12321D
SKU
LINK

Equipment
DynaMag-15
NAME
15 mL magnetic rack
TYPE
Invitrogen
BRAND
12301D
SKU
LINK

Equipment
QIAcube Connect
NAME
Automated nucleic acid extraction
TYPE
Qiagen
BRAND
9002864
SKU
LINK



Reagents:
ReagentBovilis Coronavirus Calf VaccineMerck Animal HealthCatalog #16445
ReagentHydrochloric acidSigma Aldrich
ReagentZymo DNA/RNA ShieldFisher ScientificCatalog #50-125-1706
Reagent2-MercaptoethanolSigma Aldrich
ReagentQiagen AllPrep PowerViral DNA/RNA Extraction KitQiagenCatalog #28000-50
ReagentQiagen QIAcuity One-Step Viral RT-PCR KitQiagenCatalog #1123145
ReagentGT-Digital SARS-CoV-2 Wastewater Surveillance Assay for QIAcuityGT Molecular


Consumables:
50 mL conical tubes, sterile, nuclease-free
15 mL conical tubes, sterile, nuclease-free
Ceres Nano Nanotrap Enhancement Reagent 2 (Ceres Nano, Catalog #: 10112-10)
Ceres Nano Nanotrap Magnetic Particles, in solution (Ceres Nano, Catalog#: 44202)
Qiagen Rotor Adapters (Qiagen, Catalog #: 990394)
Qiagen QIAcuity Filter Tips, 1000 microL (Qiagen, Catalog #: 990352)
PCR tubes, RNase free, low-binding
Qiagen QIAcuity Nanoplate 26k 24-well (Qiagen, Catalog #: 250002)




In addition to the materials listed above, this protocol requires access to typical microbiology equipment including pipettes, sterile, nuclease-free, filter barrier pipette tips, microcentrifuge, a Class II biosafety cabinet, a PCR cabinet, vortexes, and standard PPE.

Protocol materials
Reagent2-MercaptoethanolMerck MilliporeSigma (Sigma-Aldrich)
Materials
ReagentGT-Digital SARS-CoV-2 Wastewater Surveillance Assay for QIAcuityGT Molecular
Materials, Step 26
ReagentHydrochloric acidMerck MilliporeSigma (Sigma-Aldrich)
Materials
ReagentZymo DNA/RNA ShieldFisher ScientificCatalog #50-125-1706
Materials, Step 13
ReagentQiagen QIAcuity One-Step Viral RT-PCR KitQiagenCatalog #1123145
Materials, Step 26
Reagent2-mercaptoethanolMerck MilliporeSigma (Sigma-Aldrich)Catalog #M6250
In 2 steps
ReagentBovilis Coronavirus Calf VaccineMerck Animal HealthCatalog #16445
In Materials and 2 steps
ReagentQiagen AllPrep PowerViral DNA/RNA Extraction KitQiagenCatalog #28000-50
Materials, Step 15
Safety warnings
Prior to processing wastewater samples for the detection of the SARS-CoV-2, work with your local biosafety committee to establish biosafety protocols, and review the CDC Interim Laboratory Biosafety Guidelines for Handling and Processing Specimens Associated with Coronavirus Disease 2019 (COVID-19) available at https://www.cdc.gov/coronavirus/2019-ncov/lab/lab-biosafety-guidelines.html#environmental
Sample Collection
Sample Collection
1d
1d
Composite primary influent samples are collected (grab samples, flow composite, or time composite)
Amount50 mL sub-samples are collected in 50 mL sterile conical tubes. The samples are stored at Temperature4 °C until further processing.

Concentration of Viral Fraction
Concentration of Viral Fraction
1h
1h

Note
These steps should be performed in a Class II Biosafety Cabinet, which has been cleaned with RNase Away and water.

Thaw an aliquot of BCoVworking (ReagentBovilis Coronavirus Calf VaccineMerck Animal HealthCatalog #16445 ) TemperatureOn ice (see section Preparation of Bovilis Coronavirus). Pipette up and down to mix the aliquot and briefly centrifuge.

Add Amount62 µL of BCoVworking to each Amount50 mL wastewater sample and invert to mix.

Using a serological pipette transfer Amount10 mL of the wastewater sample into a 15 mL conical tube. Repeat this for a duplicate.

Add Amount100 µL of Nanotrap Enhancement Reagent 2 to each 10 mL wastewater sample and vortex for a few seconds.

Add Amount150 µL of Nanotrap Magnetic Virus Particles and invert 2-3 times to mix.

Incubate the samples for 10 minutes at room temperature.

Place the sample tubes in a 15 mL magnetic rack. Wait 1 minute. Remove the supernatant using a serological pipette. The magnetic particles form a pellet on the sides of the tube so it is essential to hold the tip of the pipette in the center of the tube while removing the supernatant. Transfer the supernatant to a waste container.

Add Amount1 mL molecular grade water to each 15 mL tube and remove from magnetic rack. Invert the sample to collect the Nanotrap Magnetic Particles in the water.

Transfer the sample using a pipet to a 1.5 mL tube. Place the 1.5 mL tubes on a 1.5 mL magnetic rack for 1 minute to allow the magnetic particles to separate from the supernatant. Discard the supernatant.
Add Amount200 µL ReagentZymo DNA/RNA ShieldFisher ScientificCatalog #50-125-1706 to each sample and vortex to mix. If extraction is delayed, store samples up to 3 days at Temperature4 °C .

Nucleic Acid Extraction
Nucleic Acid Extraction
2h
2h

Safety information
Proper aseptic technique should be used when working with RNA. Always wear powder-free latex, vinyl, or nitrile gloves while handling reagents, tubes, and RNA samples to prevent RNase contamination from the surface of the skin or from the environment. Change gloves frequently and keep tubes closed. During the procedure work quickly and keep everything on cold blocks when possible, to avoid degradation of RNA by endogenous or residual RNAses. Clean working surfaces and pipettes with RNase away.

The first step should be performed in the Class II Biosafety Cabinet decontaminated with UV light and RNase away. After the lysis solutions have been applied in step 9, the additional steps can take place outside of the Biosafety Cabinet.

Warm PM1 from the ReagentQiagen AllPrep PowerViral DNA/RNA Extraction KitQiagenCatalog #28000-50 at Temperature55 °C for 5-10 minutes before use.

Add Amount6 µL Reagent2-mercaptoethanolSigma AldrichCatalog #M6250 and Amount600 µL PM1 to each 1.5 mL tube containing the concentrated sample and Nanotrap Magnetic Virus Particles. Vortex to mix.



Prepare an extraction control by adding Amount6 µL Reagent2-mercaptoethanolSigma AldrichCatalog #M6250 , Amount600 µL PM1, and Amount200 µL RNA Shield to an empty tube. Also add Amount100 µL Nanotrap Enhancement Reagent 2 and Amount150 µL Nanotrap Magnetic Virus Particles.

Incubate the samples including the extraction control for 10 minutes at room temperature.



Place the 1.5 mL sample tubes on the magnetic rack and wait 1 minute before transferring the supernatant into the center tube of the rotor adapter.



Place the QIAcube rotor adapters in the QIAcube centrifuge and follow the instructions on the QIAcube control tablet to set up the shaker rack, reagents, and tips. When setting up the reagents, shake to mix the PM5 buffer. Set the elution volume to Amount100 µL . Start the extraction run.

Equipment
QIAcube Connect
NAME
Automated nucleic acid extraction
TYPE
Qiagen
BRAND
9002864
SKU
LINK


When the extraction is completed, cap the elution tubes and begin the dPCR steps or store at Temperature-80 °C if the dPCR run will occur in the following days.

Discard the used pipette tips and wipe the waste drawer and QIAcube workspace with 70% ethanol. After each run, remove the plastic tube holder and the reagent tray before running 2 cycles of UV decontamination.
Detection and Quantification of SARS-CoV-2
Detection and Quantification of SARS-CoV-2
4h
4h

Note
Technical notes:

The dPCR instrument is sensitive to microbubbles. To prevent errors in quantification and imaging, ensure that pipettes only reach the first stop when preparing this reaction.

Safety information
The preparation of the dPCR plate should be done in a PCR workstation previously decontaminated by UV light and has been cleaned with RNase away and water.

Thaw GT-Molecular controls and assay solutions on ice. If necessary, also thaw the extracted RNA on ice. Once thawed, vortex to mix.
Dilute Amount1 µL extracted RNA with Amount99 µL RNase-free water for PMMoV analysis.

Prepare master mixes for PMMoV and N1-N2-BCoV assays. Allow for one extra sample. Vortex to mix. Briefly centrifuge the tubes to collect the master mix at the bottom of the tube.
N1-N2-BCoV Master Mix ComponentsVolume/sample (microL)
Qiagen 4x One-Step Viral RT-PCR Master Mix10
Qiagen 100x Multiplex Reverse Transcription Mix0.4
GT Molecular N1-N2-BCoV Assay Solution2.0
RNase/DNase free water7.6
N1-N2-BCoV Master Mix


PMMoV Master Mix ComponentsVolume/sample (microL)
Qiagen 4x One-Step Viral RT-PCR Master Mix10
Qiagen 100x Multiplex Reverse Transcription Mix0.4
GT Molecular PMMoV Assay Solution2.0
RNase/DNase free water7.6
PMMoV Master Mix
ReagentGT-Digital SARS-CoV-2 Wastewater Surveillance Assay for QIAcuityGT Molecular
ReagentQiagen QIAcuity One-Step Viral RT-PCR KitQiagenCatalog #1123145

Pipette Amount20 µL of the appropriate master mix (N1-N2-BCoV or PMMoV) into the wells of a PCR strip tube.

Add Amount20 µL of extracted RNA sample or positive control to the PCR strip tube following the planned layout. Use the 1:100 diluted samples for the wells being used for the PMMoV assay. After transferring, pipette gently to mix. Keep the PCR strips on ice while loading.

For the non-template control: pipette Amount20 µL of molecular grade water into a PCR tube in place of adding extracted RNA.

Place a Qiagen QIAcuity 26k 24-well Nanoplate onto the Nanoplate protection tray. If the tray is not used, dust can collect on the bottom side of the plate and interfere with the imaging step. Occasionally wipe the tray with 70% ethanol to clean dust.
Using a multichannel pipette, transfer Amount39 µL of solution from the PCR strips to their respective location on the Nanoplate. Be careful to not transfer air bubbles during this step.
Note
Do not push the pipette beyond the first stop during this step.


Carefully seal the Nanoplate with a Nanoplate seal and the roller provided with the QIAcuity Instrument.
Place the sealed plate in the plate drawer inside the QIAcuity instrument.


Setup the plate by selecting "New Plate". Name the plate and choose the plate type "Nanoplate 26k 24-well".
In the dPCR Parameters section under the Priming tab, select the Qiagen Standard Priming Profile.
Under the Cycling tab create the cycling conditions shown below. These are the conditions recommended by the GT Molecular Wastewater Surveillance Guide.



In the Imaging tab create the conditions shown below. These settings are different from the GT Molecular recommended settings, but provides less frequent saturation errors.



Navigate to "Reaction mixes". Create reaction mixes named "N1-N2-BCoV Triplex Assay" and "PMMoV Assay" that contain the following details.




Navigate to "Samples and controls". Add samples names that are being quantified on this run. Extraction controls should be added as samples. Under the "Controls" tab create both a "N1-N2-BCoV Positive Control" and a "PMMoV Positive Control". Under the Non Template Controls" tab create a "N1-N2-BCoV dPCRNeg" and a "PMMoV dPCRNeg".
Navigate to "Plate Layout". Assign reaction mixes, samples and controls to their wells. Save plate and exit the setup.
On the QIAcuity tablet, select the plate and run the reaction.
Analysis and Interpretations
Analysis and Interpretations
15m
15m
When the QIAcuity run is complete, ensure the image transfer is marked as complete in the QIAcuity Software Suite before inspecting the data.
Open the plate results by selecting "Analysis". Select all the wells and targets before selecting "Show results".
In the 1D Scatterplot tab verify that the automatic threshold is accurately placed between the negative and positive partitions. If needed, adjust the threshold placement to the accurate position.
Use the "Export to CSV" button in the List tab to export the data.
Index: Preparation of Bovilis Coronavirus
Index: Preparation of Bovilis Coronavirus
1h
1h

Note
The Bovilis Coronavirus (BCoV) is used as internal process control and is added to each sample prior to filtration and extraction.

It is necessary to quantify BCoV for each new batch of BCoVworking is made from a new tube of BCoVND.

BCoV should not be vortexed at any point in the preparation or spiking of samples.

ReagentBovilis Coronavirus Calf VaccineMerck Animal HealthCatalog #16445 is lyophilized when received. Reconstitute the virus in Amount5 mL pre-chilled molecular grade water and swirl to mix.

Aliquot Amount100 µL stock in sterile 1.5 mL tubes and label each tube BCoVND (non-diluted). Store BCoVND atTemperature-80 °C

Dilute Amount60 µL BCoVND with Amount540 µL pre-chilled molecular grade water and pipette to mix. Briefly centrifuge. Label this tube BCoVINT (intermediate).

Dilute Amount500 µL BCoVINT with Amount49.5 mL pre-chilled molecular grade water and invert to mix. Aliquot Amount1 mL of BCoVworking into sterile 1.5 mL tubes and store aliquots at Temperature-80 °C .

To quantify the BCoV spike, extract the RNA of an aliquot of BCoVworking in triplicate.
Add Amount100 µL of BCoVworking working into three 1.5 mL tubes.

To each tube add Amount6 µL 2-mercaptoethanol and Amount600 µL PM1 from the Qiagen AllPrep PowerViral DNA/RNA Kit and invert to mix.

Microcentrifuge the tubes at Centrifigation13000 x g for 1 minute.

Add the supernatant to the center column of a rotor adapter and continue the extraction in the Qiagen QIAcube.
Quantify the extracted BCoVworking RNA by analyzing the extraction triplicates using the same dPCR steps beginning in the Detection and Quantification section .
To calculate BCoVworking concentration, use the average measured concentration in copies/microL of the 3 replicates analyzed by dPCR.


Index: Imaging Errors
Index: Imaging Errors
The QIAcuity instrument may show an error when viewing results that a "channel has reached saturation". To resolve this error, reduce the imaging gain and exposure duration and re-image the plate.