(i) A unidirectional workflow should be used to reduce the potential for contamination i.e. Sample preparation room (DNA extraction) > Reagent preparation room (PCR master mix) > Post-PCR room (sequencing). If you cannot use separate rooms for these steps, using separate benches and ensuring you keep the areas clean should be sufficient.
(ii) Gloves should be changed before leaving and entering each laboratory section and each time contaminating DNA is potentially encountered.
(iii) The equipment should not be moved between the sample and reagent preparation rooms.
(i) The laboratory should have a schedule for maintaining equipment including the setup, calibration, repair, record-keeping, and normal operation of all equipment. This is to verify that the equipment is functioning properly.
(ii) Temperatures of equipment (incubators, water baths, heat blocks, refrigerators, laboratory freezers) should be monitored and recorded at least once a day for each workday in use. Alternatively, temperatures may be monitored continuously using a computer-based alarm system.
(iii) The block temperature of a thermocycler should be tested at least twice a year by the laboratories or under the maintenance contract.
(iv) Use a PCR machine with a heated lid or layer/ use paraffin oil over the master mix to ensure the sample does not evaporate.
(v) The centrifuge should be balanced before use to increase bearing life and minimise vibrations that can unsettle
concentrates. Separate refrigerators and freezers for samples, reagents, and final amplification products should be maintained in the designated laboratory room.
o The positive control should produce expected bands during electrophoresis. If no amplification bands are observed in the positive control, verify the primer sequence or contact the primer supplier (assuming all other reagents are functioning properly). Use a reference strain (e.g., S. Typhi H58 or S. Typhi CT18) as a positive control for the genotyping panel. For the MDR panel, use g block amplicons for each plasmid target.
o No positive results should be detected in the negative control. If there are amplification bands present in the negative control, it indicates contamination in the primers, and the contaminated primer should be discarded, (assuming other reagents are not contaminated and working properly).
o Positive control: Use the tviD amplicon (g blocks) as the positive control.
o Negative control: Include a no-template control (nuclease-free water) with every PCR run.
(iii) Gel electrophoresis/tape station
o Before repeating check the cycling conditions on the thermocycler are correct and that none of the reagents have expired or been stored incorrectly. If there is amplification in the negative control, it indicates contamination of one or more reagents or contamination during pipetting. In this case, the samples testing positive by PCR may not be truly positive and should be repeated.
o If repeating the negative control still shows a band in the gel/Tapestation, do a thorough decontamination of all pipettes, equipment and the work space, then try using fresh reagents. When repeating the testing after deep cleaning the lab, test the negative control first to confirm as negative and then repeat the samples.
a. Bead cleanup (this applies to all the bead clean-up steps throughout the protocol)
After the final wash with ethanol, air dry the samples to remove any traces of ethanol. Any residual ethanol carryover will interfere with sequencing. Do not over-dry the samples to the point where the pellet appears cracked as it would reduce the recovery of DNA. This is because DNA fragments bind strongly to the beads and may be difficult to elute off after drying onto the beads completely.
e. Barcoding and Ligation
f. Flow cell prep and loading
o Yellow exclamation mark: The number of sequencing pores is below warranty. If this is the first time using the flow cell and it is within the warranty period (within 3 months of receipt), ONT can replace it for free, please contact your supplier for more details.
o Green tick: The number of sequencing pores is above warranty and ready for sequencing.
o Question mark: A flow cell check has not been run on the flow cell during this MinKNOW session.
When a sequencing run is complete, check the MinKNOW run report to confirm that you have enough data for analysing your results. If data is insufficient, you may start the sequencing run again for the same library already loaded on the flowcell.