Public workspacePurification of synaptic vesicles from rodent brains, troubleshooting and column cleaning/regenerating

  • 1University of Cambridge
Icon indicating open access to content
QR code linking to this content
Protocol CitationAmberley Stephens 2022. Purification of synaptic vesicles from rodent brains, troubleshooting and column cleaning/regenerating. protocols.io https://dx.doi.org/10.17504/protocols.io.4r3l28zzjl1y/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it’s working
Created: June 05, 2020
Last Modified: May 24, 2022
Protocol Integer ID: 37867
Keywords: Synaptic vesicle, vesicle purification, rat, size exclusion chromatography, gel filtration, column pouring
Abstract
Purification of synaptic vesicles from tissue samples is initially performed using sucrose gradients and ultracentrifugation. To obtain highly pure synaptic vesicle samples an additional size exclusion chromatography step is required, but can be tricky to optimise and troubleshoot. This protocol presents additional information for pouring of the chromatography column, saturating the column, cleaning the column and troubleshooting for lack of and/or loss of resolution. The purification protocol followed is from Ahmed S, Holt M, Riedel D, Jahn R. Small-scale isolation of synaptic vesicles from mammalian brain. Nat Protoc. 2013 May;8(5):998-1009
We follow the protocol provided by Ahmed, et al 'Small-scale isolation of synaptic vesicles from mammalian brain (2013)', which has proved to be a reproducible and well-designed protocol. We will only mention adaptations we made as we follow most of the protocol step by step.
We use the brains of two Sprague Dawley rats per preparation. We homogenise each brain in 9 mL of homogenization buffer and pooled the two preps at Part 1, Step 9.
For all of Part 1 and Part 2, Steps 6 and 7 we use
Equipment
Sorvall LYNX 4000
NAME
Superspeed Centrifuge
TYPE
Thermo Scientific™
BRAND
13493336
SKU

Equipment
A27-8 x 50
NAME
Fixed Angle Rotor
TYPE
Thermo Scientific™
BRAND
75003008
SKU

For Part 2 Steps 10 onwards we use
Equipment
Optima XL-90
NAME
Ultracentrifuge
TYPE
Beckman
BRAND
No longer sold
SKU

For the sucrose gradients we make the CS1 supernatant to a volume of 31 mL with homogenising buffer and layer 6 x 5 mL gradients. We use the SW41 rotor at a speed of 30,000 rpm (the ultracentrifuge can only be input in rpm). This equates to ~111,000 x g RCF av, ~154,000 x g RCF max.
Equipment
13.2 mL 14 x 89mm
NAME
Open-Top Thinwall Ultra-Clear Tube
TYPE
Beckman Coulter
BRAND
344059
SKU

Note
After a while these tubes start to crack vertically so keep an eye on them as sample will leak out.

Equipment
SW 41 Ti
NAME
Swinging-Bucket Rotor
TYPE
Beckman Coulter
BRAND
333790
SKU

Part 3. purity check by dot blotting.
This step doesn't need to be repeated once you are happy with the reproducibility of the first two parts of the protocol.
We used their suggested antibodies against Rp4
Size Exclusion Chromatography - setting up the column
Size Exclusion Chromatography - setting up the column
Part 4: size-exclusion chromatography as a final purification step.
This is the most complicated and time consuming part of the process.

The column needs to be filled at least 90% full of Sephacryl otherwise you will not get the separation between peak 1 and peak 2 containing the vesicles.
In the protocol they say you require 70 mL of packed resin, but i found i needed more.

I found the easiest way to pour the column was to use an orbital shaker and a pump (see figure)

The orbital shaker keeps the media in suspension so it is added to the column in a more homogeneous suspension - note the pumping tube needs to be stuck to the media bottle while shaking!

The pump was set to 1 mL/min which seemed a good speed to add the chromatography media. As the media is in suspended in liquid, every now and again I let it settle and opened the valve at the bottom of the column to drain the liquid to allow more chromatography media to be added.
Set up for loading sephacryl with a sample pump from a pot on an orbital shaker to keep the sephacryl in suspension

Size Exclusion Chromatography - saturating the column
Size Exclusion Chromatography - saturating the column
You will not observe vesicles coming out of the column in the first few runs, it seems the column needs to be saturated first. I emailed the authors who suggested i could try BSA if i wanted, but i felt happier just using the brain samples. It took 4 rounds of the protocol to get a protein signal, so 8 brains in total.
Size Exclusion Chromatography - peak resolution
Size Exclusion Chromatography - peak resolution
I measure the protein concentration of each fraction by hand using a nanodrop as we don't have an absorbance measurement system attached to the fraction collector. The point of sample elution changes during each run. I see two peaks, as shown in the paper. If only one peak is present either more sephacryl is needed in the column, or a smaller volume of sample should be added to the column, I try to add 200-400 uL.
As I don't have access to a method that allows counting of vesicles, I measure the protein concentration of the final sample and keep it consistent between runs. I currently get around 15 ug/uL of 50 uL vesicles. I image the vesicles using TEM to observe the size.
TEM of vesicles

To clean the sephacryl, if resolution is reducing or there is contamination. I invert the column and pump the sephacryl out into a beaker. I allow it to settle and remove the liquid. I put it on an orbital shaker and add 0.5 M NaOH and allow it to mix for a few hours before letting the sephacryl settle and pouring off the NaOH. I then rinse with water buy putting it back on the orbital shaker, then letting it settle and pouring off a couple of times. I then reload the sephacryl back into the column in 20% Ethanol.