See protocols for slicing and electrophysiological/imaging recordings for the reagents and the equipments needed in order to set up the desired experiment.
Below are indicated materials/equipment specifically required for the photostimulation protocols here described:
Wide-field LED stimulation:
Larger area photo-stimulation is available with the CoolLED pE-100 blue LED (470nm, 540mW; via Tek5 Systems in USA) with 1% intensity via remote control and 3mm light guide optical coupling to the Olympus BX-51WIF rear epi-fluorescence port. The blue light is routed to the sample with a Chroma 49002 ET eGFP large (BX2) Olympus filter cube. A manual shutter on the microscope epi-fluorescence turret is used in series to block exposure; the LED can be
remotely and synchronized activated by a TTL signal from WinFluor software and the NI-6713 output card. The minimum response timing pulse is ~1/3ms. The maximum exposure field of view with the 60x/1.0NA objective is ~415mm diameter spot; this area can be reduced by using the
condenser field iris inside the reflected fluorescence RFA module of the Olympus BX-51 upright microscope.
Targeted laser stimulation
Prairie Aurora Launch (473nm, 50mW rated laser, with ~80mW when delivered) with AOTF intensity control (sub-ms
responses) and metal clad fiber optical coupling to the Ultima scan head. A mechanical hard shutter is located within the Aurora launch (Uniblitz) to remove sample laser exposure when not photo-stimulating. The 473nm shutter is controlled by th WinFluor software via a NI-6713 card output; the control is separate in PraireView 5.2. The launch optics are designed to deliver a sub-um spot size at the focal plane of the 60x/1.0NA objective lens, but due to the 1P source there will also be lower intensity stimulation above and below the focal plane fanning out in an hourglass shape. The typical sample maximum power is ~18mW, but most active workstation systems have reduced this power to ~1.8MW maximum
with a 1.0 N.D. filter to better mimic biological responses when using sub-mm spot illumination and to move the required drive signals to a more robust region of the SIN^2 voltage-power calibration curve. Inserting the silver rod in the launch adds a lens into the blue light path which will increase the sample exposed area to ~10mm diameter at the focal plane. The Helios laser launch has the same system details for 473nm photo-stimulation, but the new launch uses the Coherent IBIS
laser system for the light source and is also designed from the block up to be much more stable with thermal changes.
Caged compounds (e.g., RUBI-GABA, RUBI-glutamate, MNI-glutamate) are dissolved in stock solutions in the dark to the desired concentration according to the manufacturer's instructions.
Caged compounds can be diluted and delivered in the recording solution, via peristaltic pump or gravity flow (depending on the preferred electrophysiology/imaging setup).
Alternatively, they can be delivered via a microinjector with the outlet positioned in immediate proximity of the recording area.