Then examine the parameters in Acqpars 'ased' mode (click 'pulse' icon), or type 'ased';
Most of the default parameters are suitable, however it's useful to compare values in the fields against those proposed by the original parameter file and the pulseprogram.
CNST4 - 1JNH value (≥100 Hz), to calculate INEPT transfer delays
CNST19 - 1HN offset (~8.0 ppm, defined in BioTop)
D1: recyle delay (~1 s for protonated samples, ~2-3 s for perdeuterated samples)
DS: 32-128 'dummy' scans that are not recorded ; allows system to reach steady state equilibration.
NS: multiple of 4; increase for increased signal to noise ( S/N increases as √NS )
-DLABEL_CN : enable 13C decoupling during 15N evolution for 13C,15N-labeled samples (default)
P1 - 15N 90º high power pulse (calibrated with BioTop)
CPDPRG2 - bi_p5m4sp_4sp.2 (adiabatic bilevel decoupling)
CPDPRG3 - 1H decoupling (garp)