May 22, 2024

Public workspaceMagnetic bead cleaning of PCR products with Cytiva Sera-Mag Select

  • 1University of Oslo
Open access
Protocol CitationAlexander Eiler, Even Garvang 2024. Magnetic bead cleaning of PCR products with Cytiva Sera-Mag Select. protocols.io https://dx.doi.org/10.17504/protocols.io.n2bvj393wlk5/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: September 13, 2023
Last Modified: May 22, 2024
Protocol Integer ID: 87715
Disclaimer
DISCLAIMER – FOR INFORMATIONAL PURPOSES ONLY; USE AT YOUR OWN RISK

The protocol content here is for informational purposes only and does not constitute legal, medical, clinical, or safety advice, or otherwise; content added to protocols.io is not peer reviewed and may not have undergone a formal approval of any kind. Information presented in this protocol should not substitute for independent professional judgment, advice, diagnosis, or treatment. Any action you take or refrain from taking using or relying upon the information presented here is strictly at your own risk. You agree that neither the Company nor any of the authors, contributors, administrators, or anyone else associated with protocols.io, can be held responsible for your use of the information contained in or linked to this protocol or any of our Sites/Apps and Services.
Abstract
Magnetic bead cleaning of PCR products, adapted for Cytiva Sera-Mag Select beads. Forked from an older protocol using AMPure XP beads with minimal changes.
Guidelines
Work at post-PCR lab.
You can use PCR tubes/strips or PCR-plates but for easiest work with magnetic beads flat-bottom 96-well microplates are recommended.
Calculate the volume needed for the beads. For normal length primers a bead:sample ratio of 0.8:1 is likely OK. For long primers you can use a ratio of down to 0.65:1 (probably even lower). If you can still see primer dimers on a gel after cleanup, try a lower bead:sample ratio.

0.8 * sample volume =
Shake the magnetic bead buffer bottle to fully resuspend magnetic particles.
Prepare 85% ethanol. You need 400 µl per sample, but prepare some extra.

Remember that when mixing ethanol and water, the total volume decreases (e.g. 15 ml water + 85 ml 100% ethanol < 100ml 85% ethanol). For the same reason, measure ethanol and water separately before mixing
Add Sample.
Add magnetic beads. Pipette mix 10 times and Incubate at room temperature for 5 minutes.
Duration00:05:00
Place the reaction plate onto a Magnet Plate for 2 minutes to separate beads from solution. (Keep on the magnet plate until step 11).
Duration00:02:00
Critical
Aspirate the supernatant from the reaction plate and discard.
Dispense 200 μL of 85% ethanol and incubate at room temperature for at least 30 seconds. Aspirate out the ethanol and discard. You can discard the ethanol by inverting the plate keeping the magnetic plate on the bottom of it.
Amount200 µL
Duration00:00:30
Repeat the washing step as above: dispense 200 μL of 85% ethanol and incubate at room temperature for at least 30 seconds. Aspirate out the ethanol and discard. You can discard the ethanol by inverting the plate keeping the magnetic plate on the bottom of it.
Amount200 µL
Duration00:00:30
Remove the remaining ethanol by pipetting.
Take the plate off the magnet
Critical
Dry off the remaining ethanol. To speed up the process, you can place the plate on a warm (30°C) heat block.
Drying takes at least 30 minutes. Check that drying was complete before continuing.
Duration00:30:00
Add 40 μL of elution buffer (PCR water or 10mM Tris), pipette mix 10 times. Make sure the buffer reaches up to the beads!
Amount40 µL
Incubate at room temperature for 2 minutes.
Duration00:02:00
Place the reaction plate onto a Magnet Plate for 1 minute to separate beads from solution.
Duration00:01:00
Critical
Transfer purified product to a new PCR plate or strips. Ensure that no beads are carried over when transferring, as it may inhibit later steps.