Then examine the parameters in Acqpars 'ased' mode (click 'pulse' icon), or type 'ased';
Most of the default parameters are suitable, however it's useful to compare values in the fields against those proposed by the original parameter file and the pulseprogram.
CNST4: 1JNH value (≥93 Hz, to calculate INEPT transfer delays, can be optimized using popt in 1D mode to maximize S/N)
D1: recyle delay (~1 s for protonated samples, ~2-3 s for perdeuterated samples
D24: 1/(4*JNH), normally set to same value as D28
DS: 32-128 'dummy' scans that are not recorded ; allows system to reach steady state equilibration.
NS: multiple of 4; increase for increased signal to noise ( S/N increases as √NS )
-DLABEL_CN : enable 13C decoupling during 15N evolution for 13C,15N-labeled samples (default)
VDLIST: J-coupling evolution delay list. Delays should be no smaller than (p21*1.273+20u)
P1 - 1H 90º high power pulse (calibrated with calibo1p1 or BioTop)
SPdB1 - power level [dB] for 1H H2O flip-back shaped pulse (calibrated with BioTop)
P3 - 13C 90º high power pulse (calibrated with BioTop)
P21 - 15N 90º high power pulse (calibrated with BioTop)