Then examine the parameters in Acqpars 'ased' mode (click 'pulse' icon), or type 'ased';
Most of the default parameters are suitable, however it's useful to compare values in the fields against those proposed by the original parameter file and the pulseprogram.
D1: recycle delay (≥2 s for protonated samples, ≥3 s for perdeuterated samples)
D21: Saturation delay. Set to >3 s (for accurate results should be >6s).
DS: 32-128 'dummy' scans that are not recorded ; allows system to reach steady state equilibration.
NS: multiple of 4; increase for increased signal to noise ( S/N increases as √NS )
-DLABEL_CN : enable 13C decoupling during 15N evolution for 13C,15N-labeled samples (default)
CNST18: 1H H2O offset in ppm (same as O1P)
CNST19: 1H HN offset in ppm (~7.8 ppm, defined in BioTop)
P1: 1H 90º high power pulse (calibrated with calibo1p1 or BioTop)
SPdB1/SPW1 - power level [dB] for 1H H2O flip-back shaped pulse (calibrated with BioTop)
P21 - 15N 90º high power pulse (calibrated with BioTop) at PLW3, high power