Public workspaceHILIC-MS analysis of central carbon metabolites in gram negative bacteria

  • 1Lawrence Berkeley National Laboratory
Icon indicating open access to content
QR code linking to this content
Protocol CitationBashar Amer, Ramu Kakumanu, Edward EK Baidoo 2022. HILIC-MS analysis of central carbon metabolites in gram negative bacteria. protocols.io https://dx.doi.org/10.17504/protocols.io.4r3l2opzxv1y/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: June 10, 2022
Last Modified: October 05, 2022
Protocol Integer ID: 64367
Keywords: Metabolomics, sugar phosphates, nucleotides, coAs, organic acids, HILIC, LC-MS
Funders Acknowledgement:
US Department of Energy
Grant ID: DE-AC02-05CH11231
Abstract
Central carbon metabolites are essential for energy metabolism, secondary metabolite and macromolecule biosynthesis in prokaryotes. Phosphorylated intermediates such as sugar phosphates are, however, difficult to resolve with conventional reversed phase (RP) chromatography. In addition, coenzyme A (CoA) compounds and nucleotide cofactors, which are also phosphorylated are difficult to resolve by RP chromatography. To this end, we have developed a hydrophilic interaction liquid chromatography quadrupole time-of-flight mass spectrometry (HILIC-QTOF-MS) method that to quantify these metabolic intermediates and cofactors in Escherichia coli DH1 and Pseudomonas putida K2440.
Attachments
Materials
Solvents used
LC-MS grade methanol (part number LC230-4, Honeywell Burdick & Jackson, Charlotte, NC, USA), LC-MS grade water (part number BJLC365-4, Honeywell Burdick & Jackson, Charlotte, NC, USA), and LC-MS grade acetonitrile (part number LC015-4, Honeywell Burdick & Jackson, Charlotte, NC, USA).

Reagents used
Concentration5 millimolar (mM) InfinityLab Deactivator Additive - Amount25 mL (part number 5191-3940, Agilent Technologies, Santa Clara, CA, USA), ammonium hydroxide (28–30% NH3 basis, part number AX1303, EMD Chemicals, Billerica, MA, USA), and ammonium acetate (part number 17836, ACS reagent, from Sigma-Aldrich, St. Louis, MO, USA).

Analytical standards
Chemical standards were purchased from Sigma-Aldrich.

MS calibrants
Reference mass correction (100 mM trifluoroacetic acid ammonium salt (part number I8720243) and 2.5 mM HP-0921 (part number 18720241)) and ESI-L-Low concentration tuning mix (part number G1969-85000) were purchased from Agilent Technologies (Santa Clara, CA, USA).

UHPLC system, LC column, and guard column
The Agilent Technologies 1290 Infinity II UHPLC system was used throughout. An Agilent Technologies InfinityLab Poroshell 120 HILIC-Z guard column (Thikness2.1 mm inner diameter, Thikness2.7 µm stationary phase particle size, part number 821725-947) was connected to an Agilent Technologies InfinityLab Poroshell 120 HILIC-Z PEEK-lined column (Thikness2.1 mm inner diameter, Thikness100 mm length, Thikness2.7 µm stationary phase particle size, part number 675775-924).

QTOF-MS system
An Agilent Technologies 6545 (part number G6545B) Quadrupole Time of Flight (QTOF) LC/MS system
was used throughout.
Safety warnings
Wear the appropriate PPE protection (i.e., gloves, safety goggles, and lab coat) and prepare solvents and LC-MS mobile phases in a chemical fume hood.
Store organic solvents in a flammable storage cabinet and peroxide-forming chemicals in the appropriate safety storage cabinets.
Sample preparation
Sample preparation
Calibration curves: Pure analytical standards were dissolved in 50:50 methanol:water (v/v) to make a stock solution of Concentration100 micromolar (µM) concentration. The stock solution was diluted 4-fold to make a standard solution of Concentration25 micromolar (µM) (with 50:50 methanol:water, v/v). A seven-point calibration curve was produced via a series of 2-fold serial dilutions (with 50:50 methanol:water, v/v), which were conducted from the Concentration25 micromolar (µM) standard solution to a standard solution ending in a concentration of Concentration0.390625 micromolar (µM) ; 2-fold serial dilutions for CoA standards, however, were conducted from a Concentration20 micromolar (µM) standard solution to standard solution ending in a concentration of Concentration0.3125 micromolar (µM) . All solutions were prepared on ice and stored at Temperature-20 °C until LC-MS data acquisition.
Extraction of intracellular metabolites: Wild type Escherichia coli DH1 (E. coli ) and Pseudomonas putida K2440 (P. putida) were grown on 0.5% glucose in LB medium at Temperature37 °C . Growth was monitored during mid exponential phase to early stationary phase. E. coli and P. putida reached average optical densities at 600 nm wavelength (OD 600 nm) of 1.67 and 2.16, respectively. Amount1.5 mL of the cell culture was harvested (in triplicate) and transferred to a Amount2 mL centrifuge tube. The cell culture was centrifuged at Centrifigation14000 x g for 2 minutes at room temperature. After removing the supernatant, metabolic activity from the cell pellet was quenched by adding 250 µL of ice-cold methanol and mixing thoroughly by vortexing for 2 minutes at Temperature4 °C . Amount250 µL of water was then added to the methanol lysate, mixed by vortexing, and centrifuged at Centrifigation14000 x g for 5 minutes at Temperature4 °C . Amount450 µL Centrifigation of the lysate was filtered via a 3k Da molecular weight cut-off centrifuge filter (Amicon-Ultra, part number UFC500396, MilliPoreSigma, St. Louis, MO, USA) at Centrifigation13000 x g at Temperature4 °C , for 60 minutes prior to LC-MS analysis.


CITATION
Baidoo EEK, Wang G, Joshua CJ, Benites VT, Keasling JD (2019). Liquid Chromatography and Mass Spectrometry Analysis of Isoprenoid Intermediates in Escherichia coli.. Methods in molecular biology (Clifton, N.J.).

Ultra high performance liquid chromatography (UHPLC) conditions
Ultra high performance liquid chromatography (UHPLC) conditions
Chromatographic separation was performed via an Agilent Technologies 1290 Infinity II UHPLC system (Table 1). The sample tray and column compartment were set to Temperature5 °C and Temperature30 °C , respectively. A sample injection volume of 1 µL was used throughout. The UHPLC mobile phases were composed of solvents (A) Concentration10 millimolar (mM) ammonium acetate, Concentration0.2 % volume ammonium hydroxide, and Concentration5 micromolar (µM) medronic acid in water and (B) Concentration10 millimolar (mM) ammonium acetate, Concentration0.2 % volume ammonium hydroxide, and Concentration5 micromolar (µM) medronic acid in Concentration90 % volume acetonitrile, and water as the remaining solvent. The mobile phases were degassed by sonication for 5 minutes. The 1290 Infinity II UHPLC system pump was purged with the aforementioned mobile phases for up to 5 minutes (at Concentration50 % volume solvent A and Concentration50 % volume solvent B) prior to LC column installation. The analytes were separated on an Agilent Technologies InfinityLab Poroshell 120 HILIC-Z PEEK lined (Thikness2.1 mm internal diameter, Thikness100 mm length, and Thikness2.7 µm stationary phase particle size) column, which was connected to an Agilent Technologies InfinityLab Poroshell 120 HILIC-Z guard column (Thikness2.1 mm internal diameter and Thikness2.7 µm stationary phase particle size). The LC column was then equilibrated with the starting mobile phase composition at a flow rate of 0.1 mL/min until the system backpressure was stable. The UHPLC gradient is described in table 2.


ABC
ComponentDescriptionPart number
1260 Infinity II Isocratic PumpIsocratic pump for internal reference lock mass deliveryG7110B
1290 Infinity II MCTColumn compartmentG7116B
1290 Infinity II MultisamplerAutosampler for automatic sample injectionG7167B
1290 Infinity II High Speed PumpUHPLC pumpG7120A
6545 Q-TOF LC/MSQuarupole-time-of-flight mass spectormeterG6545B
Table 1. The Agilent Technologies UHPLC-QTOF-MS system.



ABCD
Time (min)Percentage of mobile phase B (%B)Flow rate (mL/min)Maximum system pressure (Bar)
0900.250600
2900.250600
11.561.50.250600
11.7600.330600
12.7600.400600
12.9900.550600
14.3900.550600
Table 2. UHPLC gradient. The maximum allowable system backpressure for the column was 600 bar.




QTOF-MS method parameters
QTOF-MS method parameters
The Agilent Technologies Infinity II UHPLC system was coupled to an Agilent Technologies 6545 QTOF-MS system. The LC column effluent was delivered to the Agilent Jet Stream (AJS) ion source . An Agilent Technologies Isocratic Pump was used to deliver the reference mass correction solution. AJS facilitated the production of gas-phase [M - H]- or [M - 2H]2- ions in the negative ion mode. High mass accuracy was achieved via reference mass correction. Reference mass correction was performed with the trifluoroacetic acid (anion) [M - H]- at 112.98559 m/z (from the LC mobile phase) and the trifluoroacetic acid anion adduct of HP-0921 (i.e. hexakis(1H, 1H,3H-tetrafluoropropoxy)phosphazine (C18H18F24N3O6P3,921.23 Da, CAS NO. 58943-98-9) at 1033.98811 m/z. The reference ions were diluted to 2 µM in 80:20 acetonitrile:water (v/v) and delivered to an electrospray ionization sprayer at a flow rate of 5 µL/min via the 1260 Infinity II Isocratic Pump (Table 1). Prior to analysis, the Agilent 6545 QTOF-MS system was tuned via the ESI-L-Low calibration solution for an acquisition range of up to 1700 m/z. MS data acquisition parameters are described in table 3.


AB
QTOF-MS parametersValues
Acquisition range (m/z)70-1100 m/z
Acquisition rate (spectra/s)0.86
Nebulizer pressure (*Psi)20
Drying gas temperature (ºC)300
Drying gas flow rate (L/min)10
Sheath gas temperature (ºC)350
Sheath gas flow (L/min)12
Capillary voltage (V)3500
Fragmentor (V)100
Skimmer (V)50
OCT 1 RF Vpp (V)300
Nozzle voltage [Expt] (V)2000
Table 3. QTOF-MS parameters. *Psi is lb/in2.


LC-MS data acquisition and analysis software
LC-MS data acquisition and analysis software
LC-MS data acquisition was performed via the Agilent MassHunter Workstation software (version 8). Data processing and analysis were performed via Agilent MassHunter Qualitative Analysis (version 6) and Profinder (version 8) software.
Method validation
Method validation
Figure 1. Extracted ion chromatograms.
Figure 2. Method validation table. The mass-to-charge ratios shown are theoretical values. The CoA compounds were detected via [M - 2H]2- ions. All other compounds were detected via [M - H]- ions. Average standard and metabolite retention times are used in the last two columns of fig. 2. Phosphoglycerate data represents 2- and 3-phosphoglycerate, which could not be resolved by the method. ATP, NAD, and NADH are abbreviations for adenosine triphosphate, nicotinamide adenine dinucleotide, and nicotinamide adenine dinucleotide, reduced, respectively.

Biological data
Biological data
Figure 3. Specific sugar phosphate concentrations in E. coli and P. putida.
Figure 4. Specific CoA and nucleotide cofactor concentrations in E. coli and P. putida.

Figure 5. Specific organic acid concentrations in E. coli and P. putida.

Conclusions
Conclusions
The HILIC-QTOF-MS method was able to separate the analytes tested, except for the isomers 2- and 3-phosphoglycerate (fig. 1). The method showed good linearity for all the analytes tested, with R2 values of ≥0.99 . The limit of detection ranged from Concentration20 nanomolar (nM) to Concentration1.67 micromolar (µM) for the analytes detected. Retention time repeatability measurements were ≤ 0.49 %RSD for the analytes tested, except for pyruvate, which had a %RSD value of 1.09. There was very little difference between standard and metabolite retention times, suggesting a minimal effect of the sample matrix on retention time stability (fig. 2). The method successfully quantified intracellular sugar phosphate (fig. 3), acetyl-CoA and free CoA (fig. 4), and organic acid metabolites (fig. 5) in E. coli and P. putida metabolite extracts. Additionally, the method was able to quantify the intracellular nucleotide cofactors ATP, NAD, and NADH (fig. 4) in these microbial metabolite extracts. The metabolite concentration data suggests that the method may be applicable for other phosphorylated metabolites and organic acids.

Citations
Step 1
Baidoo EEK, Wang G, Joshua CJ, Benites VT, Keasling JD. Liquid Chromatography and Mass Spectrometry Analysis of Isoprenoid Intermediates in Escherichia coli.
https://doi.org/10.1007/978-1-4939-8757-3_11