Dec 11, 2024

Public workspaceŌGI Bio Oxygen Flask Cleaning V.2

  • 1OGI Bio
  • OGI Bio
Icon indicating open access to content
QR code linking to this content
Protocol CitationJack Hocking 2024. ŌGI Bio Oxygen Flask Cleaning. protocols.io https://dx.doi.org/10.17504/protocols.io.261ged277v47/v2Version created by Jack Hocking
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: November 02, 2023
Last Modified: December 11, 2024
Protocol Integer ID: 90317
Keywords: cleaning, OGIBio
Disclaimer
The cleaning protocols provided by OGI Bio are intended as general guidelines for the maintenance of our micro bioreactor equipment. These protocols are designed to ensure optimal performance and longevity of the equipment under standard operating conditions.
OGI Bio assumes no liability for any damage, contamination, or malfunctions resulting from deviations from these protocols, improper use, or the use of incompatible cleaning agents or methods. It is the responsibility of the user to verify that all cleaning agents and procedures are compatible with local regulations, organizational standards, and the materials of the equipment in use.
Abstract
This protocol is for the cleaning of your OGI BioReactor Oxygen Measurement Flasks.
Guidelines
If the oxygen measurement flasks are autoclaved, they should be left for 7 days before being used again and have their 100% point recalibrated prior to next use, as per our manual. They should be autoclaved at temperatures no higher than 121°C for no more than 15 minutes.
Ethanol should not be left in the flasks for more than 5 minutes.
The oxygen measurement flasks should be stored away from direct light.
Materials
70% Ethanol
Ultrapure (18 MOhm) Water


Equipment
Oxygen Measurement Flask
NAME
Accessory
TYPE
OGI Bio
BRAND
OGI-A-1008
SKU
LINK

Safety warnings
Autoclaving may damage the oxygen spots by inducing delamination of their layers. If this protocol is not suitable and you must autoclave the flasks, check the spots for visible signs of damage after autoclaving.
Due to the photosensitive nature of the spots, these flasks may not be sterilized via exposure to UV light.
Before start
Ensure protocol is carried out in sterile conditions.
Ensure you have familiarised yourself with the guidelines and warnings for this protocol.
Cleaning Oxygen Measurement Flask
Cleaning Oxygen Measurement Flask
Retaining the magnetic stir bar, empty the contents of each flask into the waste flask.
Remove stir bars and wipe thoroughly with 70% ethanol. Leave aside.
Critical
Rinse out any remaining cell culture with 10-20 mL ultrapure water. A vortex mixer can be used if there are stubborn residues. N.B. ensure that the stir bar is not in the flask if vortexing, this will destroy the spot.
Repeat Step 3 if there are still residues, dislodging with a gentle jet of water from a pipette if necessary.
Vigorously swill 10-20 mL of 70% ethanol around the flask. A vortex mixer can be used if there are stubborn residues on the flask. Take care to avoid leaving the ethanol in the flask for longer than 5 minutes.
Pour the ethanol from the flasks and leave to dry inverted, with the stir bars under the flasks. If using caps, spray them with ethanol and leave to dry inverted.
Storage
Storage
Store the flasks with their stir bars inside them with the caps tightly screwed on in a dark place. They should be left for at least 24 hours before being used again.