Sep 13, 2023

Public workspaceFluorescent Western Protocol V.2

Version 1 is forked from Fluorescent Western Protocol
This protocol is a draft, published without a DOI.
  • 1Realizing Increased Photosynthetic Efficiency (RIPE);
  • 2University of Illinois at Urbana-Champaign
Icon indicating open access to content
QR code linking to this content
Protocol CitationLynn Doran, Steven J Burgess 2023. Fluorescent Western Protocol. protocols.io https://protocols.io/view/fluorescent-western-protocol-czwux7ewVersion created by Lynn Doran
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: September 13, 2023
Last Modified: September 13, 2023
Protocol Integer ID: 87732
Keywords: Western Blot, Fluorescent Western Protocol, Protein Analysis
Abstract
Analysis of proteins using fluorescent immunoblot with optional total protein stain.

Note:
- The choice of secondary antibody depends on the choice of primary antibody, whether it is derived from a mouse (monoclonal) or a rabbit (polyclonal).
- It is advisable to stick to the 800CW wavelength to avoid problems with chlorophyll autofluorescence encountered with the 680CW antibodies.

Literature:
Guidelines
Do not touch the membrane. Always use tweezers to move the membrane to avoid protein contamination from your hands. Always use clean materials to handle and store the membranes as residual protein from dirty materials can cause high background on the membranes.
Materials
Reagents
  • InterceptR PBS Blocking Buffer (LI-COR Biosciences; 927-70001)
  • ReagentIntercept® T20 (PBS) Antibody DiluentLicorCatalog #927-75001
  • IRDye® 800CW Donkey anti-Rabbit IgG Secondary Antibody (LI-COR Biosciences; 926-32213)
  • IRDye® 800CW Donkey anti-Mouse IgG Secondary Antibody (LI-COR Biosciences; 926-32212)
  • Black Western Blot Incubation Box (LI-COR Biosciences; 929-97110)
  • 10x PBS buffer, Lonza™ BioWhittaker™ Phosphate Buffered Saline (10X) (Fisher Scientific;BW17-517Q)
  • TweenTM 20 (Fisher Biosciences; BP337-100)
  • Instant Nonfat Dry Milk
  • Methanol
  • Primary antibody (various)
  • Ultrapure water.
  • Optional: Revert 700 Total Protein Stain and Wash Solution Kit (Licor 926-11015)

Materials
  • Graduated cylinder
  • Reagent bottle
  • Black western blot incubation box (LI-COR Biosciences; 929-97301)
  • Micropipette, 1000 ul
  • Micropipette, 10 ul
  • Pipette tips, 100-1000 ul
  • Pipette tips, 1-10 ul

Equipment
  • Odyssey CLx Imager (LI-COR Biosciences)
  • Shaking platform
  • Incubator, capable of 37C.
Protocol materials
ReagentIntercept® T20 (PBS) Antibody DiluentLI-CORCatalog #927-75001
In Materials and 2 steps
ReagentIntercept (PBS) Blocking BufferLI-CORCatalog #927-70001
Step 18

Keep membranes in the black Western Blot incubation box for all steps, this is important after adding the secondary antibody because the signal is light-sensitive and will become bleached if exposed to light for a long enough period.

Note
Boxes should be thoroughly cleaned between uses. Residual protein contamination from previous blots can lead to high background signal. If background signal is increasing, clean boxes with 70% ethanol and dry thoroughly between uses.


Critical
Optional: Revert 700 Total Protein Stain
Optional: Revert 700 Total Protein Stain
2h 2m
2h 2m
Add methanol to the Revert 700 Total Protein Stain reagents as indicated on each bottle.
Rinse the transfer buffer from the membrane after transblot using ultrapure water.
Place the membrane on a Whatman 1 filter paper or drape it across the seam of the incubation box and dry at 37C for 10 minutes.

Alternatively, it can be dried overnight at room temperature if short on time on day 1.
Rehydrate the membrane for 5 minutes.
  • Nitrocellulose: rehydrate in PBS for 5 minutes with shaking.
  • PVDF: Rehydrate in 100% methanol for 30 seconds, then in PBS for 5 min with shaking.
Rinse membrane with ultrapure water.
Stain membrane with 5 mL of Revert 700 Total Protein Stain.

Measure the volume with a graduated cylinder to save expensive reagents.
Incubate with shaking for 5 minutes at room temperature.
Discard stain in appropriate hazardous waste.
Add 5 mL of Revert 700 Wash Solution.
Measure the volume with a graduated cylinder to save expensive reagents.
Incubate with shaking 30 seconds at room temperature.
Discard wash in appropriate hazardous waste.
Repeat 5 mL Revert 700 Wash.
Discard wash solution and rinse membrane with ultrapure water.

Note
This optional protocol is for total protein quantification for a single-color western blot on the 800 nm channel only. If a two-color western blot on both 700 and 800 nm channel is planned, refer to the Revert 700 Total Protein Stain Insert for the appropriate destain procedure.

Image on 700 nm channel on Odyssey Imaging System.
Note
Note: UIUC IGB reserve Odyssey Imaging System at https://www-app.igb.illinois.edu/cabbical/.


Fluorescent Western Protocol
Fluorescent Western Protocol
Wet with 1x PBS for Duration00:02:00 min

2m
Rinse membrane with dH2O
Discard PBS and incubate with ReagentIntercept (PBS) Blocking BufferLI-CORCatalog #927-70001 Duration01:00:00 at TemperatureRoom temperature
Note
Blocking prevents unspecific binding of antibody and lowers background signal.

As an alternative to Intercept Blocking Buffer you can use PBS 5% w/v milk powder (no Tween-20). Typical mini-blots use ~15 mL blocking buffer (15 mL 1X PBS, 0.75 g non-fat dry milk). Do not use milk powder as a blocking agent if protein of interest is phosphorylated.



1h
Prepare Amount1 L PBS-T solution by diluting 100 mL of 10X PBS

Incubate with primary antibody (appropriate dilution in ReagentIntercept® T20 (PBS) Antibody DiluentLI-CORCatalog #927-75001 at Temperature4 °C DurationOvernight (in cold room) with gentle agitation on a platform shaker Shaker50 rpm, 4°C .
Note
As an alternative to Intercept Antibody Diluent you can use PBST (0.1% Tween-20, 5% milk). Typical mini-blots use ~15 mL blocking buffer (15 mL 1X PBST, 0.75 g non-fat dry milk).

1:2,000 to 1:5,000 are common dilutions for primary antibodies. The ideal dilution for the primary antibody will vary based on sample type and antibody binding capacity and must be determined empirically.


Pour off the primary antibody and rinse the membrane with PBS-T.
Note
Some primary antibodies can be re-used multiple times depending on the concentration used, in this instance collect the primary antibody in a tube and store the solution at -20 oC before re-use.

Cover the membrane with PBS-T, shake vigorously on a platform shaker at Shaker50 rpm, Room temperature , 00:10:00 . Repeat 3 times.

Note
If high background signal is observed, increase wash time to 20 minutes.


Create a working dilution of secondary antibody using ReagentIntercept® T20 (PBS) Antibody DiluentLI-CORCatalog #927-75001. . For PVDF membranes only, add 0.01% SDS to the antibody diluent. Microcentrifuge secondary antibody and pipette from supernatant to precipitate out any protein complexes that may have formed during storage.

Note
As an alternative to Intercept Antibody Diluent you can use PBST (0.1% Tween-20, 5% milk). Typical mini-blots use ~15 mL blocking buffer (15 mL 1X PBST, 0.75 g non-fat dry milk).

1:20,000 is a common dilution for secondary antibodies. Consult manufacturer's recommendations and the ideal dilution for the secondary antibody will vary based on sample type and antibody binding capacity and may need to be determined empirically.

Note
Ensure that secondary antibody is compatible with primary antibody. For plants, an antibody that fluoresces at 800 nm is recommended. Chlorophyll fluoresces near 700 nm and may cause high background signal or interfere with target signal.

Incubate for Duration01:00:00 TemperatureRoom temperature with gentle agitation on a platform shaker.
Note
Incubation longer than one hour can lead to high background signal.

1h
Pour off the secondary antibody and rinse membrane with distilled water to remove residual blocking agent.
Cover the membrane with PBS-T, agitate Shaker80 rpm, Room temperature , 00:10:00
Note
If high background signal is observed, increase wash time to 20 minutes.

Discard PBS-T. Repeat step 12 three times.
Note
More washes (x5) and for longer can be done to reduce background

Rinse then cover the membrane with 1x PBS, Shaker80 rpm, Room temperature , 00:10:00 .

Proceed to imaging blot on LI-COR Odyssey CLx imaging system.
Equipment
Odyssey CLx
NAME
Imaging System
TYPE
LI-COR
BRAND
Odyssey CLx
SKU
LINK

Note
Ensure that the platform of the Odyssey CLx is thoroughly cleaned of residual protein from previous blots using isopropanol.