Column A - OutputRack → This is the rack position where the destination plates will sit. For micronic48 tube racks, these will sit in the “MTP Output Rack” position on the deck (see Software Layout figure for details).
Column B - OutputRackPosition → This is the well in the tube rack that contains the destination tube used for this step. Numbering and position of wells is the same as for the input csv file (A1 = 1, F8 = 48, A1 is at bottom left hand corner if facing the robot).
Column C - InputRack → This is the deck position where the source tubes are located. This should be the same as the rack specified in “Column B” of the Input CSV file.
Column D - InputRackPosition → This is the well location of the rack where the source tube will be located. Note that numbering and layout convention is the same as used for the Input file CSV.
Column E - DispenseWeight → The weight (in g) of biomass to be transferred from the source tube to the destination tube.
Column F - LiquidVolume → Although we NEVER use liquid dispensing on the Labman, this column must be specified on the csv file, and should always be set to 0.