| Dispose of reagent cup 1 (Pro-par) into aqueous waste. |
| Move reagent cup 2 into the 1 position. |
| Move reagent cup 3 into the 2 position. |
| Fill a clean, dry reagent cup with fresh Pro-par and place in reagent cup 3 position. |
| Dispose of reagent cup 4 (100%RA) into aqueous waste. |
| Move reagent cup 5 into the 4 position. |
| Fill a clean, dry reagent cup with fresh (100%RA) and place in reagent cup 5 position. |
| Dispose of reagent cup 6 into aqueous waste and replace with fresh 90% RA. |
| Dispose of reagent cup 33 (Define) into aqueous waste and replace with fresh Define. |
| Dispose of reagent cup 32 (Blue Buffer 8) into aqueous waste and replace with fresh Blue Buffer 8. |
| Dispose of reagent cup 31 (70% RA) into aqueous waste and replace with fresh 70% RA. |
| Dispose of reagent cup 29 (90% RA) into aqueous waste and replace with fresh 90% RA. |
| Dispose of reagent cups 16 and 17 (100% RA) into aqueous waste. |
| Move reagent cup 15 into the 17 position. |
| Move reagent cup 28 into the 16 position. |
| Move reagent cup 27 into the 15 position. |
| Fill clean, dry reagent cups with fresh 100% RA and place into the 27 and 28 position. |
| Add approximately 20 grams of DriRite to reagent cups 27 and 28. |
| Dispose of reagent cup 26 (Propar) into aqueous waste. |
| Move reagent cup 14 (Pro-par) into the 26 position. |
| Fill a clean, dry reagent cup with fresh Pro-par and place into the 14 position. |
| Dispose of reagent cup 13 (Xylene) into aqueous waste. |
| Fill a clean, dry reagent cup with fresh xylene and place into the 13 position. |
| Record maintenance performed on the maintenance sheet. |