Public workspaceExtraction and LC-MS/MS analysis of four steroids from mouse plasma and bone marrow V.1

  • 1University of Edinburgh;
  • 2Mass Spectrometry Core, Edinburgh Clinical Research Facility, University of Edinburgh;
  • 3University of Edinburgh,
Open access
Protocol CitationAndrea Lovdel, Scott G Denham, Natalie ZM Homer, William P Cawthorn 2024. Extraction and LC-MS/MS analysis of four steroids from mouse plasma and bone marrow . protocols.io https://dx.doi.org/10.17504/protocols.io.e6nvwdrmzlmk/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: February 06, 2024
Last Modified: February 20, 2024
Protocol Integer ID: 94786
Keywords: steroid, steroid analysis, quantitative steroid measurement, LC-MS/MS, corticosterone, 11-dehydrocorticosterone, testosterone, progesterone, supported liquid extraction, mouse plasma, mouse bone marrow, bone marrow, mouse, tissue specific, tandem mass spectrometry, steroid mass spectrometry
Funders Acknowledgement:
Medical Research Council
Grant ID: MR/M021394/1
NRS Research Scotland
Grant ID: EDCRF
Abstract
Steroid hormones, including progesterone, testosterone and corticosterone, play a critical role in growth, development, reproductive function and sexual differentiation. Steroid production is controlled in the adrenals via the hypothalamic-pituitary-adrenal axis and in the gonads by the hypothalamic-pituitary-gonadal axis, but steroids are also produced and metabolised in peripheral tissues such as adipose tissue. Circulating steroids (blood levels) do not always reflect local steroid tissue levels. In studies of caloric restriction there are observed changes in steroids and there is interest in the role that bone marrow adipose tissue plays1. Thus, to understand mechanisms of caloric restriction it is valuable to measure tissue levels of steroids as well as circulating levels of steroids.

Immunoassays can be used to measure steroids but they lack specificity, limited to one or two steroids, and suffer from cross-reactivity at low concentrations. Tandem mass spectrometry methods coupled with chromatographic separation are considered the gold standard analytical technique for steroid analysis, such as LC-MS/MS, with the added benefit of enabling simultaneous analysis of multiple steroids.

Here we have improved upon existing murine steroid liquid chromatography tandem mass spectrometry (LC-MS/MS) methods2 for the quantitation of four steroids in small samples to investigate the role of steroids in caloric restriction - specifically corticosterone, 11-dehydrocorticosterone, testosterone and progesterone - in bone marrow and plasma. Bone marrow is homogenised and then homogenate samples and plasma samples (~50 µL) were extracted by automated 96-well supported liquid extraction (SLE), using dichloromethane and isopropanol as an organic solvent, carried out on a Biotage Extrahera automated sample handler.

Extracted steroids were separated on a Shimadzu Nexera uHPLC with gradient elution on a Kinetex C18 column (150 x 3 mm; 2.6 µm) and a mobile phase of methanol and water (0.1% formic acid). The run time was 16 minutes, followed by mass spectral analysis on an AB Sciex 6500+ tandem quadrupole mass spectrometer operated in positive ionisation mode.

This automated SLE-LC-MS/MS method has been used to analyse 4 steroids - corticosterone, progesterone and testosterone - in mouse plasma and bone marrow. Validation demonstrates that this method is sensitive, specific, and suitable for steroid measurement in mouse bone marrow (4-8 bones) and a low volume of mouse plasma (50 µL), enabling investigation into tissue specific steroid levels and corresponding assessment of circulating steroid levels.
Guidelines
Ensure all training is up-to-date for operating the necessary laboratory instrumentation and equipment.
Materials
Consumables Table
ABCD
ItemSupplierPart no.Quantity
1.75 mL glass vials with lids Scientific Laboratory Supplies Ltd TUB1200 10
7 mL glass vials with lids Scientific Laboratory Supplies Ltd TUB1220 5
0.5 mL Microtube plasticEppendorf50
2 mL Microcentrifuge SafeLock TubesFisherBrand05-408-13850
Stainless Steel beadQiagen6998950
Glass culture tubes - 13 x 100 mm Sarstedt72.690.00196
Isolute SLE+ 400 96 well plateBiotage820-0400-P011
2 mL deep well 96 well collection plateWaters1860024821
96-well plate sealing filmVWR391-12501
Adhesive Plate SealWaters1860063361
Kinetex C18 (150 x 3 mm; 2.6 um)Phenomenex1
Kinetex KrudKatcher, 0.5 umPhenomenexAFO-84971
Deep well 96 well collection plateBiotage121-52031
Consumables for homogenisation of bone marrow and steroid extraction by supported liquid extraction (SLE)

Chemicals and Analytical Standards Table

ABC
ItemSupplierArticle no.
Water (HPLC grade) Fisher ScientificC-10449380-X
Acetonitrile (LC-MS grade) VWR 83640.320
Methanol (LC-MS grade) VWR 83638.320
Water (LC-MS grade) VWR 83645.320
IsopropanolVWR20880.320
DichloromethaneFisher ScientificC-23373320-X
CorticosteroneSigma-Aldrich/CerilliantC-107
11-dehydrocorticosteroneSteraloidsCA3690-000, powder
TestosteroneSigma-Aldrich/Cerilliant (T-037) 1 mg/mL in acetonitrile (certified)
ProgesteroneSigma-Aldrich/Cerilliant(P-069) 1 mg/mL in acetonitrile (certified)
D8-CorticosteroneSigma-Aldrich/CerilliantC-159
13C3-testosteroneSigma-Aldrich/Cerilliant(T-070) 100 ug/mL in acetonitrile (certified)
D9-progesteroneSigma-Aldrich/CerilliantP-070 100 ug/mL in acetonitrile
Formic acidFisher Scientific10596814
Ammonium hydroxideSigma-Aldrich221228
Chemicals and analytical standards

Solutions Required

  • 0.5 M ammonium hydroxide (aq) (200 mL) Make up to 200 mL with Water (HPLC grade). Mix thoroughly.

  • 98:2 Dichloromethane:Isopropanol (1 L) - Add 20 mL Isopropanol (HPLC grade) to 980 mL Dichloromethane (HPLC grade). Mix thoroughly.

  • Methanol (HPLC grade): for preparation of calibration standard/internal standard dilutions.

  • Water (HPLC grade): for preparation of calibration standards.

  • 70:30 Water:Methanol (100 mL) - Add 30 mL methanol (LC-MS grade) to 70 mL water (LC-MS grade). Mix thoroughly.


Equipment Table
ABC
ItemModelSupplier
Dri-blockDB.3A Techne
Microtube centrifuge Heraeus Freso 21 D-37520 LR56495 Sigma
TissueLyser II for cell lysisIIQiagen
Liquid Chromatography Pump LC30ADShimadzu
Autosampler SIL-30ACMPShimadzu
Column oven , Nexera X2CTO-20ACShimadzu
QTrap 6500+ mass spectrometer5038125-JAB Sciex
Gilson RepetmanGilson RepetmanGilson
Analytical BalancePS-100Fisher Scientific
Deepwell plate thermoshakerTS-DWGrant Scientific
Liquid handling robotExtraheraBiotage, Sweden
SPE Dry 96 dual evaporatorSPE DryBiotage, Sweden
Equipment required for homogenisation, extraction and steroid analysis


Safety warnings
Attention
Ensure risk assessments are up to date and that all local laboratory guidelines are followed for handling chemicals and biological samples
Ethics statement
All mouse studies were approved by the University of Edinburgh Animal Welfare and Ethical Review Board and were done under project licenses granted by the UK Home Office.
Before start
Ensure all consumables are in stock and all compounds and reagents are freshly prepared
Bone Marrow Collection and Homogenisation
Bone Marrow Collection and Homogenisation
1h 41m 30s
Prepare 0.5 mL Eppendorf microtubes for all bone marrow samples by removing the bottom of a 0.5 mL microtube (Eppendorf) with a razor blade
Batch sizes of 20-40 bone marrow samples are most manageable. Isolate bone marrow (BM) from frozen (or fresh) tibia by doing the following:
Remove tibiae from freezer and transfer frozen tibiae from dry ice onto wet ice prior to cutting.
Use a razor blade to cut off the proximal and distal ends of each tibia
Place each cut tibia into a cut-off 0.5 mL microtube, to keep the bone upright, and add this tibia in a 0.5 mL microtube into a 2 mL microtube.
Centrifuge the 2 mL cap-lock microtube and its contents in a microcentrifuge forCentrifigation8000 rcf, 4°C, 00:01:00 . The BM pellet will collect at the base of the 2 mL microtube

1m
Record the mass of each BM pellet, using an analytical balance, prior to homogenisation. Multiple BM pellets can be combined at this point. (e.g. 2, 4 , 6 or 8 BM pellets can be homogenised together).
Add Amount500 µL acetonitrile w/ 0.01% formic acid to the bone marrow in the 2 mL microtube

Add a 5 mm stainless steel metal bead into a 2 mL microtube, cap securely and homogenise in the Qiagen bead mill homogeniser at 30 Hz for Duration00:00:30

30s
Remove and transfer the homogenisation tube containing the homogenate to the freezer for Duration01:00:00 at Temperature-20 °C

1h
Remove the homogenate tube from the freezer, centrifuge for Duration00:05:00 at 16,100 rcf

5m
Transfer the homogenate to a clean, labelled glass vial using a glass pipette
Add another aliquot of Amount500 µL acetonitrile w/ 0.01% formic acid to the bone marrow pellet, vortex and centrifuge for Duration00:05:00 at 16,100 rcf

5m
Transfer the supernatant from the homogenisation tube and add it to the first supernatant in the glass tube.
Reduced the supernatant to dryness under nitrogen at Temperature40 °C

30m
Resuspend the dried down homogenate supernatant ofSampleSample by adding Amount200 µL water prior to preparation of calibration standards (4), working internal standard (5) and then supported liquid extraction (See point 7) of steroids

Preparation of mouse plasma for extraction
Preparation of mouse plasma for extraction
Remove mouse plasma samples from the freezer and defrost on ice. Label up glass culture tubes with mouse sample IDs and aliquot Amount50 µL of each SampleSample into labelled glass tubes, alongside bone marrow. Record exact volume of plasma aliquoted in plate map.

Preparation of calibration standard solutions
Preparation of calibration standard solutions
Prepare 100 µg/mL stock solutions of each steroid - corticosterone (B), 11-dehydrocorticosterone (A), testosterone (T) and progesterone (P4) in 1.75 m
Prepare a mixed stock of the 4 steroids - B, A, T, P4 - by using 100 µg/mL stock solutions. Do this by adding 50 µL x 100 µg/mL B, 50 µL x 100 µg/mL A, 50 µL x 100 µg/mL T and 50 µL x 100 µg/mL P4 + 800 µL methanol to give a 5 µg/mL stock.
Dilute the 5 µg/mL stock Mixed STOCK by 1:10 dilution (100 µL x 5 µg/mL + 900 µL methanol ) to give 500 ng/mL stock
Dilute the 500 ng/mL mixed STOCK by 1:10 dilution (100 µL x 500 ng/mL + 900 µL methanol ) to give 50 ng/mL stock
Dilute the 50 ng/mL mixed STOCK by 1:10 dilution (100 µL x 5 µg/mL + 900 µL methanol ) to give 5 ng/mL stock
Dilute the 5 ng/mL Mixed STOCK by 1:10 dilution (100 µL x 5 µg/mL + 900 µL methanol ) to give 500 pg/mL stock
Dilute the 500 pg/mL Mixed STOCK by 1:10 dilution (Amount100 µL x 5 µg/mL + Amount900 µL methanol ) to give 50 pg/mL stock Go togo to step #6

Preparation of internal standard solution
Preparation of internal standard solution
Prepare 100 µg/mL solutions of each isotopically labelled internal standard (d8-corticosterone, 13C3-testosterone and d9-progesterone) in methanol.
Prepare a mixed 5 µg/mL Internal Standard mix stock solution of the three isotopically labelled steroids by adding Amount25 µL x 100 µg/mL d8-corticosterone, Amount25 µL x 100 µg/mL 13C3-testosterone andAmount25 µL x 100 µg/mL d9-progesterone toAmount425 µL methanol.

Prepare a 5 ng/mL Working Internal Standard solution by taking 10 µL x 5 µg/mL Int Std Mix + 1990 µL methanol.
Preparation of calibration standards
Preparation of calibration standards
Prepare calibration standards directly into labelled glass culture tubes using the following table for volumes of each stock concentration, into a final volume of 200 µL water.
ABCD
Standard name Amount (ng) STD Mix Vol (uL)Vol water (uL)
0 STD 0 0 200
0.00250 STD 0.00250 5 uL x 500 pg/mL 195
0.00500 STD 0.00500 10 uL x 500 pg/mL 190
0.01000 STD 0.0100 20 uL x 500 pg/mL 180
0.0250 STD 0.0250 5 uL x 5 ng/mL 195
0.0500 STD 0.0500 10 uL x 5 ng/mL 190
0.100 STD 0.100 20 uL x 5 ng/mL 180
0.250 STD 0.250 5 uL x 50 ng/mL 195
0.500 STD 0.500 10 uL x 50 ng/mL 190
1.00 STD 1.00 20 uL x 50 ng/mL 180
2.50 STD 2.50 5 uL x 500 ng/mL 195
5.00 STD 5.00 10 uL x 500 ng/mL 190
10.0 STD 10.0 20 uL x 500 ng/mL 180
Calibration standard preparation table
Supported liquid extraction of steroids from calibration standards and samples
Supported liquid extraction of steroids from calibration standards and samples
10m
Transfer 200 µL of prepared calibration standards, bone marrow homogenate and mouse plasma from glass culture tubes into a 2 mL deep well 96-well collection plate (Biotage), prepared in Microsoft Excel template (see Files) and following a plate map design as below -
Plate Map - Column-wise plate layout for automated Supported Liquid Extraction on an Extrahera liquid handling robot (Biotage, Sweden)

Using a multi-step pipette enrich the plate containing calibration standards with IS by adding Amount20 µL x 5 ng/mL Working Internal Standard into each calibration standard, including 0 std and each sample (bone homogenate and mouse plasma), except for the Double blank and solvent blank.

Using the Extrahera liquid handling robot, set up with an SLE400 extraction plate and a deep well extraction plate, aliquot 200 µL Concentration0.5 Molarity (M) ammonium hydroxide in water into each well of the 96-well plate.

Transfer Amount400 µL of liquid from each well (containing sample and the diluent, Concentration0.5 Molarity (M) ammonium hydroxide) into a 400 µL volume Supported Liquid Extraction plate (SLE400), pre-placed into the deck on the Extrahera, with a deep well Waters 2 mL deep well collection plate below, pre-labelled with the batch details and date of extraction.

Allow the diluted sample to adsorb onto the SLE extraction bed for Duration00:05:00 before eluting with Amount600 µL x 98:2 (v/v) dichloromethane/isopropanol and repeating twice more, each time collecting the eluent into the collection plate

5m
Dry down the eluent collected into the 2 mL collection plate using the SPE Dry down for 96-well plates under nitrogen.
Resuspend in Amount100 µL x 70:30 water/methanol, seal the plate with a zone-free plate seal and shake on ThermoShaker for Duration00:05:00 at Shaker300 rpm

5m
Place the plate in the autosampler for LC-MS/MS or store atTemperature-20 °C until ready for analysis.

Steroid analysis by LC-MS/MS
Steroid analysis by LC-MS/MS
16m
Set up an acquisition batch in Analyst software using the electronic file of the calibration standards and sample list. Set to inject 10 µL per sample and use a method of chromatographic separation as described in step 16 and 17 and mass spectrometer settings as outlined in steps 18 and 19.
Set up the Shimadzu Nexera X2 liquid chromatography system and fit with a Phenomenex Krud Katcher and a Phenomenex 150 x 3 mm; 2.6 µm Kinetex C18 liquid chromatography column, using mobile phase A - water with 0.1% formic acid and mobile phase B - methanol with 0.1% formic acid at 0.5 mL/min and Temperature40 °C .

Set up chromatographic gradient as below with a run time ofDuration00:16:00 per sample
ABCD
Time (min) Flow (mL/min)A (%) B (%)
Initial 0.5 45 55
4.00 0.5 45 55
10.00 0.5 0 100
12.00 0.5 0 100
12.10 0.5 45 55
16.00 0.5 45 55
Chromatographic gradient details. A - water w/ 0.1% formic acid; B - methanol w/ 0.1% formic acid. 40oC. Kinetex C18 (150 x 3 mm; 2.6 µm)

16m
Set up the mass spectrometer for Multiple Reaction Monitoring (MRM) method in positive mode, with electrospray ionisation as below, with divert of LC flow into the mass spectrometer set at 1 minute and 12 minutes.
AB
Instrument Sciex QTrap 6500+
Source, Ionisation Mode IonDrive Turbo V Source, ESI
Scan Mode, Polarity MRM, Positive
Resolution (Q1/Q3) unit/unit
Mass range Low mass
Pause Time 5.007 ms
Acquisition time 16.0 min
Delay time 0 sec
Curtain Gas (CUR) (N2) 30 units
Collision Gas (CAD) (N2) Medium
IonSpray Voltage (IS) (Positive) 4500 V
Temperature (TEM) 600 °C
Ion Source Gas 1 (GS1) (Air) 40 units
Ion Source Gas 2 (GS2) (Air) 60 units
Entrance Potential (EP) (Positive) 10 V
Probe position (x – axis) 5
Probe position (y – axis) 2
Mass Spectrometry source settings for positive ion electrospray ionsiation on QTrap 6500+

Set up the mass spectrometer to monitor for the following MRM transitions for each steroid and isotopically labelled steroid.

ABCDEFG
Q1 Mass (Da) Q3 Mass (Da) Scan time (msec) Steroid NameDP (V)CE (V)CXP (V)
347.1 121.1 50Corticosterone 1 76 29 8
347.1 90.9 50 Corticosterone 2 76 75 12
345.1 121.2 50 11-Dehydrocorticosterone 1 66 31 12
345.1 91.2 50 11-Dehydrocorticosterone 2 66 83 40
289.1 97.0 50Testosterone 1 101 29 12
289.1 109.250 Testosterone 2 101 31 6
315.1 97.1 50 Progesterone 1 96 23 10
315.1 109.1 50 Progesterone 2 96 27 10
355.3 128.1 50 d8B-Corticosterone 1 37 45 14
355.3 125.0 50 d8B-Corticosterone 2 2956 14
292.1 100.2 50 13C3-Testosterone 101 29 12
324.1 100 50 d9-Progesterone 96 23 10
MRM settings for each steroid, including quantitative (1) and qualitative (2) ions for each steroid. DP - declustering potential, CE - collision energy, CXP - collision exit potential

Check the retention times of the steroids are as expected, as shown in the chromatogram below:
Expected result
Retention times; 11-dehydrocorticosterone at 4.1 mins, corticosterone at 6.2 mins, testosterone at 8 mins and progesterone at 9 mins


Overlaid Ion Chromatogram of MRM transitions for 11-dehydrocorticosterone, corticosterone, testosterone and progesterone, separated on a Kinetex C18 (150 x 3 mm; 2.6 um) at 0.5 mL/min


Once the chromatography has been checked and the retention times are consistent, set the batch of samples to analyse. Use MultiQuant software and excel to evaluated the LC-MS/MS data to calculate the concentration of steroids in each sample, as detailed:
Protocol
Using MultiQuant and Excel software to evaluate and report multi-analyte targeted LC-MS/MS data
NAME
Using MultiQuant and Excel software to evaluate and report multi-analyte targeted LC-MS/MS data
CREATED BY
Natalie ZM Homer
.
Protocol references
1.Cawthorn WP, Scheller EL, Parlee SD, Pham HA, Learman BS, Redshaw CM, Sulston RJ, Burr AA, Das AK, Simon BR, Mori H, Bree AJ, Schell B, Krishnan V, MacDougald OA. Expansion of Bone Marrow Adipose Tissue During Caloric Restriction Is Associated With Increased Circulating Glucocorticoids and Not With Hypoleptinemia. Endocrinology. 2016 Feb;157(2):508-21. doi: 10.1210/en.2015-1477. Epub 2015 Dec 22. PMID: 26696121; PMCID: PMC4733126.

2. Verma M, Sooy K, Just G, Nixon M, Morgan R, Andrew R, Chapman KE, Homer NZM. Quantitative analysis of 11-dehydrocorticosterone and corticosterone for preclinical studies by liquid chromatography/triple quadrupole mass spectrometry. Rapid Commun Mass Spectrom. 2020 Sep;34 Suppl 4(Suppl 4):e8610. doi: 10.1002/rcm.8610. Epub 2020 Feb 11. PMID: 31677354; PMCID: PMC7540072.