Oct 04, 2023

Public workspaceEx vivo mouse brain patch clamp recordings and Fura-2 imaging

  • 1Northwestern University
Open access
Protocol Citationtaylor.panczyk 2023. Ex vivo mouse brain patch clamp recordings and Fura-2 imaging. protocols.io https://dx.doi.org/10.17504/protocols.io.eq2lyj8zrlx9/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: October 03, 2023
Last Modified: May 31, 2024
Protocol Integer ID: 88723
Keywords: ASAPCRN
Abstract
In this protocol we detail the steps to perform ex-vivo brain slices electrophysiology and Fura-2 recordings. For both types of recordings, the brain slices with the desired nucleus must be placed into the recording chamber and patch clamp high resistance seal must be performed. For cell attached configuration recording, the sual must be kept unbroken. For whole cell configuration, quick pulses of negative pressure to break into the cell. For Tura-2 recordings, a whole cell patch clamp must be performed to allow the Fura-2 to enter the cell.
Materials

Solutions to prepare:

Internal solutions:
These are prepared prior to the experiment day and aliquoted in 1.5 ml tubes, and stored at -20°C until the day of experiment.
K-based internal solution for whole cell
Compound M.W. Conc (mM)
KMeSO4 150.20 126.00
EGTA 380.40 1.00
KCl 74.55 14.00
HEPES 238.30 10.00
Mg-ATP 551.10 2.00
Na3-GTP 523.20 0.50
CaCl2 110.98 0.50
MgCl2 203.32 3
Adjust pH to 7.3 with KOH
Internal solution for Fura 2
Compound M.W. Conc (mM)
KMeSO4 150.20 135.00
EGTA 380.40 5.00
KCl 74.55 5.00
Na-PCr 255.10 10.00
HEPES 238.30 5.00
Mg-ATP 551.10 2.00
Na3-GTP 523.20 0.50
CaCl2 110.98 0.50
Adjust pH to 7.3 with KOH

Electrophysiology solutions:
These are prepared as 10´ stock solutions (below) and diluted for use on the day of the experiment.
Stock aCSF
Solution 10x         1L
Compound MW g conc (mM)
NaCl 58.44 79.33 1357.5
KCl 74.55 1.86 25
sodium bicarbonate 84.01 21 250
sodium phosphate 120 1.5 12.5
Physiological glucose

aCSF 1x                 1L
ABCD
Compound MW g conc (mM)
NaCl diluted from 10x 125
KCl diluted from 10x 2.5
sodium bicarbonate diluted from 10x 25
sodium phosphate diluted from 10x 1.25
CaCl2 1M solution 2ml 2
MgCl2 1M solution 1ml 1
Glucose* 0.63 3.5
* add glucose to working solution (do not include in 10x stock)


High glucose aCSF
Solution 1x         1L
ABCD
Compound MW g conc (mM)
NaCl diluted from 10x 135.75
KCl diluted from 10x 2.5
sodium bicarbonate diluted from 10x 25
sodium phosphate diluted from 10x 1.25
CaCl2 1M solution 2ml 2
MgCl2 1M solution 1ml 1
Glucose* 4.5 25
* add glucose to working solution (do not include in 10x stock)

Prepare patch pipettes
Prepare patch pipettes
Turn on the Sutter P-1000 puller and enter the desired pull protocol.

Insert a thick-walled borosilicate glass capillary and press pull.
Pipette resistance must be of 2.5 to 5 megaohms.
Setting up patch rig and environment
Setting up patch rig and environment
Turn on the MultiClamp 700B Amplifier, Axon Digidata 1550B digitizer, micromanipulator, computer tower and the associated software. Note: amplifier and digitizer must be turned on prior to opening software.
Turn on O2/CO2 tank and bubble aCSF solution.
Take an aliquot of internal solution from the -20° fridge.

Fill syringe with internal solution (or aCSF for cell attached recordings), place a filter on the end of the syringe, and place a MicroFil Pipette Filler on the end of the filter.
Turn on the pump and circulate recording aCSF solution through chamber.

Adjust and set the rate of the pump to 3-4 mL/min.

Turn on water heater and set to desired temperature (~Temperature34 °C )

Examine slices and patching cells
Examine slices and patching cells
Transfer brain slice from incubation beaker to the recording chamber.
Secure down slice with a harp (slice anchor).
Locate and focus the desired brain region under the 4x objective.

Change the microscope lens to the 60x objective.
Slowly focus on healthy neurons in slices for patching.
For whole cell configuration, fill a glass micropipette one-third full of internal solution. For cell attached configuration fill it with aCSF. Ensure there is no residual internal solution on exterior of glass micropipette, as this may introduce salts into the micromanipulator and add additional noise to recordings. Remove any air bubbles by gently flicking the glass micropipette.
Gently place the glass micropipette onto the wire electrode and tighten.
Position the electrode using a micromanipulator.
Under the 60x objective, bring the tip of the glass pipette above the slice.
Apply a positive pressure and maintain it.
Approach the cell diagonally. The positive pressure should create a small dimple on the cell.
Once a dimple is formed, zero the pipette voltage, release the positive pressure, and apply a small amount of negative pressure. The resistance should begin to increase rapidly.
As the resistance increases, clamp the cell at your resting potential of interest (typically - 60 mV).
After a giga-ohm seal is formed, leave it this way to record in cell attached configuration or apply a few quick pulses of negative pressure to break into the cell to record in whole cell configuration.
Start recording.
Fura 2 imaging
Fura 2 imaging
Transfer slices to a 2PLSM and turn on the 780-nm wavelength laser.
Add Fura-2 (100 uM) and Alexa Fluor 594 hydrazide (25 uM) to the internal solution.
Fill the patch pipettes with the internal solution.
Perform a whole cell patch clamp recording as previously explained.
Wait 10 minutes until the cell is completely filled with the internal solution containing Fura-2.
Choose a dendrite from the patched neuron 20-100 um far from soma.
Draw a strait line ROI over the chosen dendrite.
Start line scan acquisitions with 0.195-m pixels and 12-s dwell time.