Public workspaceDetection of avian Influenza Virus H5N1 with a Real time PCR kit

  • 1Genekam Biotechnology AG
Open access
Protocol CitationSudhir Bhatia, Gudrun Baersch 2024. Detection of avian Influenza Virus H5N1 with a Real time PCR kit. protocols.io https://dx.doi.org/10.17504/protocols.io.5qpvok38bl4o/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: May 24, 2024
Last Modified: May 28, 2024
Protocol Integer ID: 100540
Keywords: Avian influenza Virus H5N1, Real time Polymerase chain reaction, cow milk, poultry, zoonosis, wild life, livestock, duck, Virology
Disclaimer
Genekam Biotechnolgy AG
Duissernstrasse 65a
47058 Duisburg
Germany
Abstract
Avian Influenza virus H5N1 has caused a number of outbreaks in human beings in recent years. Recently it has caused outbreak in human and cows in USA. Therefore it is essential to detect the presence of this virus as early as possible in order to monitor the presence of this virus, hence we have developed a number of ready to use PCR kits to detect this virus. Highly pathogenic strains are reponsible for such outbreak. It spreads also in poultry with very fatal consequences. The kit double- and triple-checks the presence of this virus, therefore it makes possible to detect the H5N1 strains circulating in different countries worldwide, moreover mutated versions are likely to be detected with double and triple check method. The cross reaction panel has done with related and non related pathogens like Influenza virus strains: H3N2, H1N1, H7N3, HIN1 (California strain); RSV, Enterovirus, CMV, Influenza B, Mycoplasma etc.
Materials
FR006 - FluHunter: Avian Influenza Virus H5N1 (Realtime PCR-Kit) CE
-UDI-DI: 04262420430102

Compositions of the Kit:

- Tube A (2 tubes): 2 x 385µl
- Tube B (2 tubes): 2 x 550µl
- Tube Y (1 tube): 110µl
- Positive (+Ve) Control (tube D1) (1 tube): 40µl
- Negative (-Ve) Control (tube D2) (1 tube): 60µl
- Realtime PCR thermocycler
- Laboratory centrifuge
- 96 well microplates for PCR or Microtubes (0.2ml)
- Sterile Pipette-tips with filter (1-10µl & 10-100µl)
- Pipettes (quality pipettes)
- Vortexer
- RNA isolation kit: Genekam Biotechnology AG, Germany or equivalent
Safety warnings
Attention
Precautions:
- It is only for in vitro use.
- It must be used through trained persons.
- It should not be used after expiry date.
- During PCR, one should work very cleanly.
All works must be done in different parts of laboratory.
- NEVER STORE THIS KIT WITH FOOD, VACCINE OR BACTERAL CULTURES.
- Decontaminate the instruments regularly (once a week).
- To dispose of the kit and its contents, add 70% ethanol to the contents of the kit and then dispose of it.
Before start
1. User must read the manual before use. 2. Important Hint:
Real time PCR is based on fluorogenic dyes. There are 2 probes with 2 different dyes in this kit hence you have to program them in your machine:

- First the probe, namely Carboxy-fluorescein (reporter) and 6-Carboxy tetramethyl rhodamine (quencher). The results will be shown as Ct -Values. Up to 40 Ct should be taken positive. Value between 40-45 Ct should be taken as marginal positive (doubtful).
- Second the probe, namely HEX (reporter, it is available as VIC in some machines) and 6-Carboxy tetramethyl rhodamine (quencher). The results will be shown as Ct -Values. Up to 40 Ct should be taken positive. Value between 40-45 Ct should be taken as marginal positive (doubtful).
If any of the both channels or both are positive, it is to be considered positive.
Thaw one tube each: A, B, Y, D1 and D2. After thaw, put the tubes at 4°C (as it is better). If the kit is not in use, store them at -20 ° C.
Mark your microtubes with a sample number,  +ve Control and –ve Control. Otherwise use a 96 well microwell plate instead of tubes.
Add 7µl of Tube A to each tube.
Add 10µl of B to each micro tube. Avoid touching the wall of the microtubes.
Add 1µl of Y to each tube (Avoid touching the wall of the microtubes).
TIP: User can calculate the total requirement of chemicals needed. User can mix 7µl of A + 10µl of B + 1µl Y together in one tube for one reaction, but to have 10 reactions, there will be total volume of 180µl (70µl Tube A, 100µl B and 10µl Y). From this, 18µl can be distributed into each tube. This step saves time and hardware.
Add 2µl of your RNA with sterile pipette-tip with filter to each micro tube according to your label except +Ve and -Ve (Avoid touching the wall). Use a new pipette tip for each sample!
Use new pipette tip with filter. Add 2µl of Tube D1. This is the positive control supplied with our kit.
Use a new pipette tip. Add 2µl of –Ve (tube D2) to –Ve Control (Do not touch the wall).
Centrifuge all tubes at 8000 rpm for 5 sec. (This is not necessary, but it is better).
Run the program of your thermocycler as follows: Check if everything is added correctly, as the volume of each microtube must be almost the same.
Now enter quencher and reporter dye to setup your software (see FAQ) and run the following program (2h30min):
1.         60 minutes at 42°C             10 minutes at 70°C 2.         15 seconds at 95°C                x 45 cycles             60 seconds at 60°C
Example of setting up a real time machine: Instrument Setup Instructions for real time PCR (RT-PCR System: ABI 7500)
Open SDS Software -File New -Select: Absolute Quantification assay -Insert name -Click: Next



Create a New Detector 1: Select Reporter and Quencher-Dye: (FAM-TAMRA) -Click: Add -Select: Passive Reference-none -Click: Next



Create a New Detector 2 (VIC-TAMRA) -Click: Add -Select: Passive Reference-none -Click: Next



Insert Well positions: Unknown – or -NTC (No Template Control) - or– Standard -Click: Finish: The Plate Document will open.



Write your Array: Name of Samples for each well or tube.



Select Instrument Tab – Insert the Program into the Thermal Profile as mentioned in point 11 - Change Sample Volume (20µl) -Click: File - “Save as” to save the Setup -Click: “Start” to begin the PCR run: Shortly the run time will appear on the screen.
Before starting the PCR program, check that the plate or tubes are properly sealed. The wells of the well plate or microtubes must be in contact with metal block (important!). There should be no air or lose contact with metal block of thermocycler. In case of 96-well plate, it should be sealed with adhesive cover. Now run your PCR.
Once the PCR is finished, a small dialog box will appear: “The run has been completed successfully”
Click: OK. Now you can take out the microtubes or well plate.
STEP B
Click the Result-Tab and the Amplification-Tab: Place the Threshold line above the background, then select: “Analyse” (see Figure below). Calculate the threshold cycle (Ct) Value for each well. There should be no signal in the negative control. Successful positive controls or positive samples must give a curve in the software graphics.



The Report page will give you the Ct-values and Standard deviation for each well.
If either channel is positive, the result should be considered positive regardless of whether the internal control is negative or positive. If the pathogen-specific test and the internal control are negative, the test should be repeated with the same sample or a new isolation. If the internal control is positive and the H5N1 test is negative, the test should be scored negative.