Public workspaceDeposition of matrix using an M5 TM sprayer for high resolution MALDI analysis V.3

  • 1Vanderbilt University
Icon indicating open access to content
QR code linking to this content
Protocol CitationMartin Dufresne, Angela R.S. Kruse, Jamie Allen, Danielle Gutierrez, Jeff Spraggins 2023. Deposition of matrix using an M5 TM sprayer for high resolution MALDI analysis. protocols.io https://dx.doi.org/10.17504/protocols.io.yxmvmnzmbg3p/v3Version created by Angela R.S. Kruse
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: April 05, 2023
Last Modified: October 18, 2023
Protocol Integer ID: 80059
Abstract
This protocol describes the use of an HTX M5 TM sprayer to deposit small molecule matrices onto tissue sections for high resolution MALDI imaging mass spectrometry (IMS). Spraying conditions will vary depending on matrix, and this protocol focuses on matrices for lipid IMS.
Safety warnings
All steps involving organic solvents should be performed in a fume hood.
Autofluorescence Scan
Autofluorescence Scan
Remove slides from freezer and place in desiccator for Duration00:30:00

30m
Collect autofluorescence scan using a Zeiss AxioScan Z1: Pre-IMS Autofluorescence Microscopy
TM Sprayer Setup
TM Sprayer Setup
Set nitrogen flow to 8 psi. turn on TM sprayer.
Open HTX software
Set nozzle temperature to 60C, and heated stage temperature (if applicable) to 45C.
Clean sample loop with 10mL of methanol with sprayer set to "load". Switch to "spray" and allow to spray for 2 minutes.
Make matrix solution at a concentration of 12mg/mL in a solution of 60% THF and 40% acetone containing 5uL/mL DMF. Sonicate solution for 10 minutes.

Note that the matrix concentration and solvent composition may vary by sample and matrix.
Switch sprayer to "load" and load sample loop with 2 mL of matrix solution using a syringe.
Switch sprayer to "spray" and allow to spray for ~30 seconds. Check that matrix is spraying by placing a glass slide under the nozzle.
Tape slide onto heated stage, and adjust the spray area in the HTX software.
Set the method in the HTX software:
1350 mm/min nozzle velocity
1.5 mm track spacing
0.05 mL/min flow rate
CC pattern
4 passes
40 mm nozzle height
No drying time
Ensure that the method is highlighted. Press start under the Cycle tab. Click Continue. Wait for purge delay and confirm that the nozzle is moving
When matrix deposition is finished, remove the slides and store them in a slide box. Press "valve load confirm" in the software.
Cleanup
Cleanup
Set TM sprayer nozzle and stage temperatures to 30C
With sprayer set to "load", load loop with 10 mL each of acetonitrile followed by methanol. Set sprayer to "spray" and allow to flow for ~5 minutes.
Spray nozzle tip with methanol, taking care not to touch the tip.
Clean the stage with methanol. Do not spray solvents directly onto the stage.
Replace the bench liner in front of the instruments
Dispose of syringes in biohazard container
Shut down HTX software and close the computer. Turn off the TM sprayer.
Shut off the nitrogen flow.