Public workspaceC-SOP-201: Genomic DNA Quantification using a Qubit Fluorometer

  • 1The Centre for Genomic Pathogen Surveillance, Oxford, United Kingdom
Open access
Protocol CitationMihir Kekre, Ben Pascoe 2023. C-SOP-201: Genomic DNA Quantification using a Qubit Fluorometer. protocols.io https://dx.doi.org/10.17504/protocols.io.kqdg3xb47g25/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: November 06, 2023
Last Modified: November 20, 2023
Protocol Integer ID: 90468
Keywords: Qubit, DNA quantification, DNA QC, Quantus, fluorometric quantification, dna quantitation
Funders Acknowledgement:
National Institute for Health and care Research
Grant ID: 16/136/111
Disclaimer
This protocol has been adapted from dsDNA quantification assays developed by ThermoFisher Scientific.
Abstract
In order to prepare high quality double-stranded DNA libraries for Illumina WGS, extracted genomic DNA needs to be accurately quantified. Qubit Fluorometers detect fluorescent dyes that are specific to the target of interest. These fluorescent dyes emit only when bound to specific target molecules, even at low concentrations. This provides obvious advantages over UV absorbance quantitation, which measures anything absorbing at 260 nm — DNA, RNA, protein, free nucleotides, or excess salts. Moreover, UV spectrophotometry often does not have the sensitivity to accurately measure low concentrations of DNA and RNA. The Qubit dsDNA broad-range (BR) and high-sensitivity (HS) assays are highly selective for double-stranded DNA (dsDNA) over RNA and single-stranded DNA (ssDNA).

This protocol has been adapted from dsDNA quantification assays developed by ThermoFisher Scientific.

Variations of this method include quantification assays from Promega:


This is an open-access protocol distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike.





Guidelines
Storage (Qubit BR/HR kit):
When not in use, the buffer and BR/HS reagent (dye) should be stored at TemperatureRoom temperature while the standards must be stored at Temperature4 °C .
The reagent is light and humidity sensitive, it should be stored in the dark and exposed minimally during assay and sample preparation. When stored as directed, the reagent has a shelf-life of about 6 months.
Materials
1. Qubit Assay Tubes (ThermoFisher Scientific, Cat no. Q32856)

2. Qubit dsDNA BR (Broad Range) or High Sensitivity (HS) Assay Kit (Thermo Fisher Scientific, Cat no. Q32853 or Q32854)

3. Single or multi-channel pipettes (P2, P10, P200 and P1000) and compatible sterile, filtered tips

4. 5.0 or 15.0 ml Falcon tubes (BD Falcon, 5 ml: Cat no. 352196, 15 ml: Cat no. 352070 or equivalent)

5. Qubit 2.0/3.0/4.0/Flex Fluorometer
(Different versions of the Qubit fluorometer may be used - check that the kit is compatible with your version. Set up instructions may differ slightly)

6. Vortex

7. Microcentrifuge

Safety warnings
Attention
Take proper precautions and wear appropriate PPE when handling potentially hazardous chemicals. Ensure that chemicals, spent containers, and unused contents are disposed of in accordance with local safety standards.


Qubit dsDNA Broad Range (BR) Reagent Kit: Refer to ThermoFisher Scientific's SDSs for additional information.
• BR Reagent 200x in DMSO: GHS Category 4 for flammability, no data is available on the mutagenicity or toxicity of the reagent, it is known to bind nucleic acid so the same safety precautions should be used as when handling other potential mutagens.

• BR Buffer: No GHS classification but may cause irritation with susceptible persons.
Before starting
Before starting
The buffer and BR/HS reagent (dye) should be stored at TemperatureRoom temperature while the standards must be stored at Temperature4 °C .

The reagent is light and humidity sensitive, it should be stored in the dark and exposed minimally during assay and sample preparation. When stored as directed, the reagent has a shelf-life of about 6 months.
Prior to initiating the protocol, ensure that all active workbenches are cleaned with 80% ethanol, all relevant personal protective clothing is worn and the work area is prepared for DNA quantification according to local GLP guidelines.

Create an organised bench space by clearing away all clutter in order to maximize work efficiency.
Qubit Mastermix and Standard/Sample Preparation
Qubit Mastermix and Standard/Sample Preparation
Set up the required number of Qubit tubes for standards and samples.

Note
i. The Qubit dsDNA assays require two standards (#1 and #2) for calibration.

ii. If you have already performed a calibration on the Qubit machine for the selected assay you can use the previous calibration stored on the machine. It is recommended to perform a new calibration for every sample batch but a same-day calibration would be fine to use for multiple batches.

Label the sample assay tube lids with sample IDs and the standard tubes lids as S1 and S2.
Note
i. Use only thin-wall, clear, 0.5mL PCR tubes. The recommended consumables are listed under Materials (Cat. No. Q32856).

ii. Do not label the side of the tube as this could interfere with the sample reading

Prepare a Qubit Mastermix for each sample and standard by diluting the Reagent in a 1:200 ratio with Buffer.

Per sample:
Qubit® dsDNA BR/HS Reagent: Amount1 µL
Qubit® dsDNA BR/HS Buffer: Amount199 µL

To avoid pipetting errors, it is recommended to add some overage as indicated below in Table 1.

ABCDEF
Volume per sample (µL)For 10 samples (x 11)For 20 samples (x 21)For 30 samples (x 31)For N samples
Qubit Reagent1112131N+1
Qubit Buffer199218941796169199 x (N+1)
Table 1

Note
To avoid any cross-contamination, it is recommended that the total amount of working solution required for samples and standards is removed from the working solution bottle and then added to the mastermix container instead of pipetting directly from the stock bottle to each tube.

Pipetting
Mix
Critical
Aliquot the appropriate volume of Qubit Mastermix into each tube:
  • Standard tubes: Amount190 µL
  • Sample tubes Amount198 µL


Next, to the appropriate tube:
  • Add Amount10 µL of standards (S1 and S2).
  • Add Amount2 µL of each sample.


The final volume in each tube must be Amount200 µL once the samples/standards are added.
Note
i. The Qubit dsDNA assays require 2 standards for every round of calibration.

ii. For samples that are suspected to have low yields of DNA, the sample tube volume can be decreased to anywhere between 180-198µL in order to supplement the final volume with a greater quantity of input DNA.

iii. When adding 1–2µL of sample input, use a P-2 pipette for best results.



Pipetting
Mix
Critical
Cap each tube and then mix vigorously by vortexing for Duration00:00:05 .

Allow all tubes to incubate at TemperatureRoom temperature for Duration00:02:00 .

2m 5s
Incubation
Reading Standards and Samples with the Qubit 4.0
Reading Standards and Samples with the Qubit 4.0
On the home screen of the Qubit, press dsDNA, then select dsDNA: High Sensitivity or dsDNA: Broad Range depending on the kit being used.


Computational step
INSTRUMENT CALIBRATION:
If calibration is being performed, press Read Standards to proceed. (If a calibration is not necessary, you can omit steps 8.1 to 8.3).




Insert the tube containing Standard #1 into the sample chamber, close the lid, and press Read Standard. When reading is complete, remove the tube and place back in the rack.
Insert the tube containing Standard #2 into the sample chamber, close the lid, then press Read Standard. When reading is complete, remove the tube and place back in the rack.
The instrument displays the results of the calibration immediately after S1 and S2 have been read. For further information on interpretation of the results, refer to the Qubit 4.0 Fluorometer User Guide.
Next, press Read Samples to proceed to the assay.


On the Sample Volume screen, select the volume of sample added to the tube and the output concentration units (i.e. ng/μl or other).
Insert the first sample tube into the sample chamber, close the lid, and then press Read Tube.

Note
i. The software displays the results of each sample. The top value (in large font) is the concentration of the original sample. The bottom value is the dilution concentration. Record the concentration of the original sample, remove the tube and repeat steps 9.1 and 9.2 to record results for each additional sample.

ii. To manage and import assay readings directly from the instrument, refer to the relevant Fluorometer User Guide in the next section.

Analyze