Keywords: Qubit, DNA quantification, DNA QC, Quantus, fluorometric quantification, dna quantitation
Funders Acknowledgement:
National Institute for Health and care Research
Grant ID: 16/136/111
Disclaimer
This protocol has been adapted from dsDNA quantification assays developed by ThermoFisher Scientific.
Abstract
In order to prepare high quality double-stranded DNA libraries for Illumina WGS, extracted genomic DNA needs to be accurately quantified. Qubit Fluorometers detect fluorescent dyes that are specific to the target of interest. These fluorescent dyes emit only when bound to specific target molecules, even at low concentrations. This provides obvious advantages over UV absorbance quantitation, which measures anything absorbing at 260 nm — DNA, RNA, protein, free nucleotides, or excess salts. Moreover, UV spectrophotometry often does not have the sensitivity to accurately measure low concentrations of DNA and RNA. The Qubit dsDNA broad-range (BR) and high-sensitivity (HS) assays are highly selective for double-stranded DNA (dsDNA) over RNA and single-stranded DNA (ssDNA).
This protocol has been adapted from dsDNA quantification assays developed by ThermoFisher Scientific.
Variations of this method include quantification assays from Promega:
When not in use, the buffer and BR/HS reagent (dye) should be stored at Room temperature while the standards must be stored at 4 °C.
The reagent is light and humidity sensitive, it should be stored in the dark and exposed minimally during assay and sample preparation. When stored as directed, the reagent has a shelf-life of about 6 months.
2. Qubit dsDNA BR (Broad Range) or High Sensitivity (HS) Assay Kit (Thermo Fisher Scientific, Cat no. Q32853 or Q32854)
3. Single or multi-channel pipettes (P2, P10, P200 and P1000) and compatible sterile, filtered tips
4. 5.0 or 15.0 ml Falcon tubes (BD Falcon, 5 ml: Cat no. 352196, 15 ml: Cat no. 352070 or equivalent)
5. Qubit 2.0/3.0/4.0/Flex Fluorometer
(Different versions of the Qubit fluorometer may be used - check that the kit is compatible with your version. Set up instructions may differ slightly)
6. Vortex
7. Microcentrifuge
Safety warnings
Take proper precautions and wear appropriate PPE when handling potentially hazardous chemicals. Ensure that chemicals, spent containers, and unused contents are disposed of in accordance with local safety standards.
Qubit dsDNA Broad Range (BR) Reagent Kit: Refer to ThermoFisher Scientific's SDSs for additional information.
• BR Reagent 200x in DMSO: GHS Category 4 for flammability, no data is available on the mutagenicity or toxicity of the reagent, it is known to bind nucleic acid so the same safety precautions should be used as when handling other potential mutagens.
• BR Buffer: No GHS classification but may cause irritation with susceptible persons.
Before starting
Before starting
The buffer and BR/HS reagent (dye) should be stored at Room temperature while the standards must be stored at 4 °C.
The reagent is light and humidity sensitive, it should be stored in the dark and exposed minimally during assay and sample preparation. When stored as directed, the reagent has a shelf-life of about 6 months.
Prior to initiating the protocol, ensure that all active workbenches are cleaned with 80% ethanol, all relevant personal protective clothing is worn and the work area is prepared for DNA quantification according to local GLP guidelines.
Create an organised bench space by clearing away all clutter in order to maximize work efficiency.
Qubit Mastermix and Standard/Sample Preparation
Qubit Mastermix and Standard/Sample Preparation
Set up the required number of Qubit tubes for standards and samples.
Label the sample assay tube lids with sample IDs and the standard tubes lids as S1 and S2.
Prepare a Qubit Mastermix for each sample and standard by diluting the Reagent in a 1:200 ratio with Buffer.
Per sample:
Qubit® dsDNA BR/HS Reagent: 1 µL
Qubit® dsDNA BR/HS Buffer: 199 µL
To avoid pipetting errors, it is recommended to add some overage as indicated below in Table 1.
A
B
C
D
E
F
Volume per sample (µL)
For 10 samples (x 11)
For 20 samples (x 21)
For 30 samples (x 31)
For N samples
Qubit Reagent
1
11
21
31
N+1
Qubit Buffer
199
2189
4179
6169
199 x (N+1)
Table 1
Aliquot the appropriate volume of Qubit Mastermix into each tube:
Standard tubes: 190 µL
Sample tubes 198 µL
Next, to the appropriate tube:
Add 10 µL of standards (S1 and S2).
Add 2 µL of each sample.
The final volume in each tube must be 200 µL once the samples/standards are added.
Cap each tube and then mix vigorously by vortexing for 00:00:05.
Allow all tubes to incubate at Room temperature for 00:02:00.
2m 5s
Reading Standards and Samples with the Qubit 4.0
Reading Standards and Samples with the Qubit 4.0
On the home screen of the Qubit, press dsDNA, then select dsDNA: High Sensitivity or dsDNA: Broad Range depending on the kit being used.
INSTRUMENT CALIBRATION:
If calibration is being performed, press Read Standards to proceed. (If a calibration is not necessary, you can omit steps 8.1 to 8.3).
Insert the tube containing Standard #1into the sample chamber, close the lid, and press Read Standard. When reading is complete, remove the tube and place back in the rack.
Insert the tube containing Standard #2 into the sample chamber, close the lid, then press Read Standard. When reading is complete, remove the tube and place back in the rack.
The instrument displays the results of the calibration immediately after S1 and S2 have been read. For further information on interpretation of the results, refer to the Qubit 4.0 Fluorometer User Guide.
Next, press Read Samples to proceed to the assay.
On the Sample Volume screen, select the volume of sample added to the tube and the output concentration units (i.e. ng/μl or other).
Insert the first sample tube into the sample chamber, close the lid, and then press Read Tube.
Additional Information & Troubleshooting
Additional Information & Troubleshooting
For more information, refer to the following links depending on the model of instrument and type of assay being used.