Then examine the parameters in Acqpars 'ased' mode (click 'pulse' icon), or type 'ased';
Most of the default parameters are suitable, however it's useful to compare values in the fields against those proposed by the original parameter file and the pulseprogram.
D1: recyle delay (~1 s for protonated samples, ~2-3 s for perdeuterated samples)
DS: 32-128 'dummy' scans that are not recorded ; allows system to reach steady state equilibration.
NS: multiple of 4; increase for increased signal to noise ( S/N increases as √NS )
CNST21 - 13CO offset (~173 ppm, same as O1P, defined in BioTop)
CNST22 - 13CA offset (~48 ppm, defined in BioTop)
CPDPRG2 - 1H decoupling (garp)
P21 - 15N 90º high power pulse (calibrated with BioTop)
CPDPRG3 - 15N decoupling (garp, garp4)