Calibrate mass and mobility: Adjust funnel pressure by selecting the lightning bolt icon in timsControl software (Display mobilogram in [V].
Adjust funnel pressure until the calibrant is at the appropriate [V].
Positive mode: m/z 622.029 should have 132.5 [V]
Negative mode: m/z 601.979 should have 118.5 [V]
Repeat mobility calibration.
Wait 20 minutes to assess drift. If system has not drifted in pressure, do not recalibrate. If system has drifted in pressure, recalibrate.