Aug 16, 2022

Public workspaceBiological sample lysis and extraction using the Covaris LE220+ for LC-MS

  • 1University of Manchester
Icon indicating open access to content
QR code linking to this content
Protocol Citationronan o'cualain 2022. Biological sample lysis and extraction using the Covaris LE220+ for LC-MS. protocols.io https://dx.doi.org/10.17504/protocols.io.8epv598b6g1b/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: July 20, 2022
Last Modified: August 16, 2022
Protocol Integer ID: 67105
Keywords: Protein extraction, Covaris LE220+, Sample lysis, biological, clinical, sample preparation, LC-MS, ultrasonication, mass spectrometry, proteomics, microproteomics, tissue, cells, sonication
Abstract
This protocol details the procedure of sample lysis and protein extraction using the Covaris LE220+ ultrasonicator.
Attachments
Guidelines
Initial assumptions and preparation:

  • Allow approximately Duration00:30:00 for sample lysis and protein extraction.
  • You may pre-book the use of the LE220+ machine beforehand on PPMS.
  • If it is your first time using the LE220+, let us know at bioms@manchester.ac.uk, you will be provided with training, in which you will be provided with login rights to the machine and some basic processing methods to get you started.
  • You have cells or tissue in Covaris tubes, TemperatureOn ice .
Materials
Locate the following buffers, consumables, and reagents:
AB
Location Buffer/reagent
Fridge 02 N/A
Bench 2x stock of S-trap lysis buffer (10% SDS in 100 mM TEAB, pH 7.5) LC-MS grade water Eppendorf tubes
Freezer N/A

ReagentSDSSigmaCatalog #75746
ReagentTriethylammonium bicarbonate (TEAB)Sigma AldrichCatalog #T7408





Identify the following equipment that you will use:

  • p20 pipette, p200 pipette, p1000 pipette and pipette tips
  • Stainless steel rack for LE220+ processing (500282 rack for 520045 tubes, 500452 rack for 520185 tubes) – Ask a member of BioMS staff for it.
  • Spare Covaris tubes for making up the rows in the rack to either 8 per row for the 500282 rack, or 4 for the 520185 rack, check next to LE220+ machine.




Before start

Initial preparation:

Before you begin:
Log onto the LE220+ and check that the “run” box in the top right hand corner is green. If not, it may be that the system is not at temperature. Click the “System status” box on the right hand side – if a red “X” is next to temperature, then you must wait until the system reaches the temperature that you have set the system to in your method before you begin. The default temperature for the system is Temperature10 °C .


Preparing samples and LE220+
Preparing samples and LE220+
You will need to dilute the S-Trap lysis buffer to a working stock concentration.
Note
S-trap lysis buffer (10% SDS) is provided in the orange trays on bench.





Using the 1mL pipette, add Amount500 µL of 2x stock of S-Trap lysis buffer to Amount500 µL of LCMS grade water.
This is the working stock concentration (5% SDS, Concentration50 millimolar (mM) TEAB Ph7.5 ).
Note
If you know that your samples are high in disulphide bonds, you may add Concentration5 millimolar (mM) of dithiothreitol (DTT) at this stage. Refer to the section on “reduction and alkylation” for instructions on how to prepare DTT.

Pipetting
Add S-Trap lysis buffer (5% SDS with Concentration50 millimolar (mM) TEAB Ph7.5 ).
Note
Do not freeze your samples once they contain buffer, it will cause the glass tube to crack.

Pipetting
If using the 520045 tubes, add approximately Amount50 µL to Amount130 µL of buffer, the final volume will depend on the amount of sample already present in the tube.

Note
It is not recommended by Covaris to use Amount100 µL in 520045 tubes, the headspace caused by the liquid to air ratio impacts on the performance of the instrument

Pipetting
If using the 520185 tubes, add up to Amount500 µL of buffer, again this is dependent on the amount of sample that you wish to process.
Pipetting
Add the Covaris tubes to the respective stainless steel rack.
Note
According to Covaris, all spaces in the row must be filled, there are spare tubes beside the machine for this purpose. Tubes are processed in rows on the width of the plate.

Covaris tubes in their racks - NB: sample tubes are processed in rows of either 4 or 8, with no gaps.


Clamp the lid on the rack finger-tight, and keep the samples in the rack TemperatureOn ice until you are ready for processing.





Sonolab 8.2 software
Sonolab 8.2 software
On initial training, we will provide you with a general method that should be sufficient to lyse your sample. At the end of the process, if you feel that your sample would benefit from more treatment, simply run the program again.


Note
Log onto the LE220+ focussed ultrasonicator and check that the “run” box in the top right hand corner is green. Sonolab software controls the LE220+ system.
LE220+ focussed ultrasonicator Sonolab software.
If not, it may be that the system is not at temperature. Click the “System status” box on the right hand side.
Settings
– if a red “X” is next to temperature, then you must wait until the system reaches the temperature that you have set the system to in your method before you begin. The default temperature for the system is Temperature10 °C .

How will you know if your sample is ready?

With cell suspensions, the sample will turn from opaque to transparent, a sign that the cells have lysed.
With tissue, after a quick centrifuge spin, you will see that your pellet is less than before.
Centrifigation
Sonolab will automatically select the last method run by the user.
To select an alternative method, use the drop-down menu to load an alternative method, it is located on the HOME tab, in the top left corner.
To specify the rows of the plate that you wish to process, click “edit”.
This will open up the edit parameters page.
If you have two rows of tubes to process, click on the process in the left hand side.
Note
This will allow the rows on the plate display on the right hand side of the page to be selected.

Click on the two rows that you wish to process.
Note
When the LE220+ begins, it will process one row, then the next, in that order. The total processing time for what you specify in this page is displayed in the bottom right hand corner.

Click save and return to the main menu.
The Run method will remain grey and inactive until a method is selected and all necessary conditions are met.
On the right hand side, in the “Home” tab, take a look at the “System status” tab – if it is shaded red and contains a red X, then a starting parameter(s) has not been achieved – click on the button to see the issue.
Note
Usually it is because the set temperature has not been reached. By default the LE220+ is set to Temperature10 °C .


If your method is set to Temperature6 °C , simply wait until the temperature has been reached.


Running your method
Running your method
Click on the “Motion Control” box, and click on “load”, the plate holder will move to the front of the LE220+ for plate loading.
Open the door of the LE220+ by simultaneously pressing the green door lock button while gently pulling the handle of the safety enclosure Perspex door.
Load the plate with the orientation that you specified in the method.

Note



When the “Run” button is green, the selected method may be started. Click the “Run” button.
Note
As the method runs, the graphic for the row under treatment flashes green, and the actual average incident power and temperature data are displayed.

When the method has finished, the message “Method Complete” is displayed. Click “OK”, then press the green door lock button and open the safety enclosure door. Remove the treated sample plate and place TemperatureOn ice .
Label Amount0.5 mL or Amount1.5 mL tubes for each 520045 or 520185 tube that you are processing, and transfer the lysed sample extract to the labelled tube.
Note
You may now freeze your samples at this stage for storage, or proceed to reduction and alkylation.



If finished, logout of the Sonolab software and return the rack to the side of the LE220+ for the next user.
Note
IMPORTANT: The Sonolab software should be always left ON, as by shutting it down, the default settings are triggered and the day to day maintenance settings are lost. If the software needs to be re-booted, contact a member of the facility.

Do not shut down the software!

Notes:

Treatments are defined by the following parameters: (For an in-depth explanation of these parameters, go to covaris.com)

  • Treatment time.
  • Cycles per Burst – the number of acoustic oscillations contained in each burst. The illustration below shows five cycles in each burst.
  • Duty Factor – the percentage of active burst time in the acoustic treatment. The illustration shows a Duty Factor of 20%.
  • Peak Incident Power – the power, in Watts, being emitted from the transducer during each burst.
  • Average Incident Power - Peak Incident Power multiplied by the Duty Factor.
(For an in-depth explanation of these parameters, go to covaris.com)