License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: In development
We are still developing and optimizing this protocol
Created: May 22, 2024
Last Modified: February 19, 2025
Protocol Integer ID: 100290
Keywords: 3D hCANCO
Funders Acknowledgements:
National Science Foundation
Grant ID: 1946970
Abstract
Purpose
Three-dimensional hCANCO dipolar correlation experiment at moderate spinning rate.
Scope
Inter-residue (CAi, Ni, COi-1) backbone chemical shift correlations of CA, N and CO atoms. 3D hCANCO is used for sequential assignment of 15N and 13C chemical shifts in conjugation with 3D hNCACX and hNCOCX.
Checklist for setting hCANCO3d experiment:
Load parameters file or previous standard experiment acquired using the same spectrometer, probe, and spinning rate (if exist).
Set carrier frequencies for each channel (O1p, O2p, O3p, cnst16, cnst17).
Set acquisition time (aq), dwell time (dw), spectral width (sw), and complex points (TD) for each dimension.
Set recycle delay (d1)
Set NUS schedule if applicable
Set scan numbers (NS)
Guidelines
This SOP is written based on the pulse sequence developed at NMRFAM. If using other pulse sequences, the strategy is still applicable, but the parameter names will be different. Please refer to the schematic pulse sequence at section 6.10 to find corresponding parameter names if using other pulse sequences
Materials
Definitions:
Term
Definition
hCANCO
1H-13C CP based CANCO 3D correlation experiment
MAS
Magic Angle Spinning
CP
Cross Polarization
NUS
Non-Uniform Sampling
Instrument: The hCANCO3d example dataset in this SOP was acquired using a 600 MHz NMR spectrometer with Bruker Avance III console. The probe is a Phoenix 1.6 mm HCN probe. For the best results, this experiment should be executed after properly adjusting the shimming and magic angle.
Sample:
Type: Protein
Labeling: Fully protonated U-13C,15N labeling (default), or 13C-glycerol skip-labeled (alternative conditions where noted). Not applicable to perdeuterated samples.
Minimum amount: 100 nmol (3.2 mm rotor) or 50 nmol (1.6 mm rotor)
hCANCO3d schematic pulse sequence:
Example Parameter set for a Phoenix 1.6 mm probe using 13.333 kHz MAS on 600 MHz Bruker spectrometer (pulse sequence: CNC3d-NAN):
Operator: User should be knowledgeable to operate MAS probes and use Topspin. User should know the power limit of the MAS probe being used.
Before start
MAS rate: 13.333 kHz at 600 MHz, 16.667 kHz at 750 MHz, or 20 kHz at 900 MHz (~88.9 ppm in 13C).
Temperature: As determined for optimal sample sensitivity and resolution.
Below optimizations are required before setting up the hCANCO3d experiment (parameters needs to be updated are shown in parenthesis). Example SOP for each optimization is attached at References. Note that rf powers in this pulse sequence defined using kHz rather than Watt (i.e., input kHz number for rf powers and the code will calculate the corresponding Watt numbers for Topspin to use). Only the hard pulses for calibration need to be input as Watt (plw13, plw14, and plw15).
Calibration of 1H, 13C, and 15N solid pulses (p14, plw14, p13, plw13, p15, and plw15)
Optimization of HCA CP (p43, spnam41, cnst41, spnam31, and cnst31)
Optimization of CAN CP (p35, spnam32, cnst32, spnam52, and cnst52)
Optimization of NCO CP (p53, spnam33, cnst33, spnam53, and cnst53)
Optimization of CA selective pulse (p38, spnam38, cnst38)
Optimization of high-power 1H decoupling (cpdprg4, p24, and cnst24)
Setup time: ~15 min, presuming initial calibrations in the cited SOPs have already been completed.
Procedure
Procedure
Load pulse program and parameter set. “CNC3d-NAN” pulse program and parameter set “hCANCO3d-NAN_par”.
If a previous hCANCO3d experiment acquired using the same setup is available (i.e., same spectrometer,
probe, pulse program and MAS rate), open the previous dataset and type “edc”. Input the directory for the new experiment and press “OK”.
If no previous data available, then use Topspin commend “edc” to open a new experiment. Input “CNC3d-NAN” as PULPROG. Type “rpar”, then load “hCANCO3d-NAN_par”.
Set the 1H carrier frequency (O2p) at ~5 ppm, 13C carrier frequency (O1p) at 175 ppm, and 15N carrier frequency (O3p) at 118 ppm. For example, type “O1p” and enter, then input 175 to the opened window.
Optimize the pulse sequence parameters. The optimization needs to be done are listed in the "Before start" section under "Guidelines & Warnings". Example SOPs for the required optimization are attached at "References". Note that the parameters can be optimized use different methods/pulse sequences. Please consult to your facility manager to perform the optimization.
Set the carrier frequencies of 13C channel.
Set cnst16 to 175 ppm. This constant is used as the frequency for detection and 3rd NCO CP. Type “cnst16”, then input 175.
Set cnst17 to 55 ppm. This constant is used as the frequency for the 1st hCA CP, 2nd CAN CP, and the soft selective pulse for CA. Type “cnst17”, then input 55.
Set the acquisition parameters.
Direct dimension (t3 or F3)
Default acquisition time: depending on 13C T2, default is ~20ms (alternative for 13C-glycerol: ~30 ms).
Default spectral width: ~300 ppm.
Default carrier frequency: ~175 ppm.
Indirect 13C dimension (t1 or F1)
Increment of delay (IN_F, s): n * τr. For a given MAS rate, adjust the increment of delay to integer multiples of τr that covers at least 30 ppm 13C spectral width.
Spectral width (SW, ppm): coupled with increment of delay, >30 ppm is required to cover the full spectrum
Carrier frequency (O1p. ppm): 175 ppm.
Hypercomplex scheme (FnMODE): States-TPPI
Maximum evolution time (AQ, s): depending on 13C T2, the default is ~10 ms. The evolution time equals to (Increment of delay) * (TD). Once Increment of time is fixed as described above, change number of TD to adjust the total evolution time.
Indirect 15N dimension (t2 or F2)
Increment of delay (IN_F, s): n * τr. For example, 375 ms can be used for 13.33 kHz MAS on 600 MHz. For a given MAS rate, adjust the Increment of delay to integer multiples of τr that covers at least 40 ppm 15N spectral width.
Spectral width (SW, ppm): coupled with increment of delay, >40 ppm is required to cover the full spectrum.
Carrier frequency (O3p, ppm): 118 ppm.
Hypercomplex scheme (FnMODE): States-TPPI
Maximum evolution time (AQ, s): depending on 15N T2, the default is ~10 ms. The evolution time equals to (Increment of delay) * (TD). Once increment of delay is fixed as described above, change number of TD to adjust the total evolution time.
Set the recycle delay
Use recycle delay of 1.3*T1 for maximum sensitivity per unit time. If 1H T1 is not measured, use the default value which is 2 s. However, measuring 1H T1 before setting this experiment is highly recommended.
Due to high-power 1H decoupling is applied, don’t set recycle delay less than 1s.
Set NUS schedule. This step is optional, but 25% NUS is recommended for 3D experiments.
Adjust measurement time as required by incrementing number of scans in multiples of 16.
Validation
Start the experiment and monitor the first ~20-30 rows
Process first dimension FT to check for adequate signal correctly arraying indirect dimension.
Protocol references
Reference: J. Pauli, M. Baldus, B. van Rossum, H. de Groot, H. Oschkinat, Backbone and side-chain 13C
and 15N signal assignments of the aspectrin SH3 domain by magic angle spinning solid-state NMR at 17.6 Tesla, ChemBioChem 2 (2001) 272–281. https://doi.org/10.1002/1439-7633(20010401)2:4<272::AID-CBIC272>3.0.CO;2-2
XX Y. Li, D.A. Berthold, H.L. Frericks, R.B. Gennis, C.M. Rienstra, Partial 13C and 15N chemical-shift assignments of the disulfide-bond-forming enzyme DsbB by 3D magic-angle spinning NMR spectroscopy, ChemBioChem 8 (2007) 434–442. https://doi.org/10.1002/cbic.200600484