Apr 22, 2022

Public workspaceBackflush of Dead-end Ultrafilter

  • 1USFDA
Icon indicating open access to content
QR code linking to this content
Protocol CitationAndrea Ottesen, Brandon Kocurek 2022. Backflush of Dead-end Ultrafilter. protocols.io https://dx.doi.org/10.17504/protocols.io.14egnzp6yg5d/v1
Manuscript citation:
Microbiota of the Hickey Run Tributary of the Anacostia River Robin Cagle , Padmini Ramachandran , Elizabeth Reed , Seth Commichaux , Mark K. Mammel , David W. Lacher , Daniela Miller , Sabina Lindley , Mihai Pop , Michael C. Bazaco , Beverly Wolpert , Errol A. Strain , Scott Aker , Joan Feely , and Andrea Ottesen Julia A. Maresca , Editor
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it’s working
Created: November 06, 2021
Last Modified: March 22, 2023
Protocol Integer ID: 54903
Disclaimer
DISCLAIMER – FOR INFORMATIONAL PURPOSES ONLY; USE AT YOUR OWN RISK

The protocol content here is for informational purposes only and does not constitute legal, medical, clinical, or safety advice, or otherwise; content added to protocols.io is not peer reviewed and may not have undergone a formal approval of any kind. Information presented in this protocol should not substitute for independent professional judgment, advice, diagnosis, or treatment. Any action you take or refrain from taking using or relying upon the information presented here is strictly at your own risk. You agree that neither the Company nor any of the authors, contributors, administrators, or anyone else associated with protocols.io, can be held responsible for your use of the information contained in or linked to this protocol or any of our Sites/Apps and Services.
Abstract
The backflush procedure pushes out all the cells and particles that were captured during the pumping procedure. It is most easily carried out in the lab with established (pump holding equipment) or ad hoc installation of pump, flushing material , filter and a >500 sterile vessel to collect the backflush.
Image Attribution
Andrea Ottesen
Guidelines
Ultrafilter is backflushed with a Amount500 mL solution of 0.5% Tween 80

Materials
  • Geotech Geopump Peristaltic Pump
  • Kimwipes/paper towels
  • Scissors
  • Bench protector paper, to catch water from leaks (Fisher Scientific, Cat. 14-206-65, or similar)
  • Faceshield (Fisher Scientific, Cat. 17-310, or similar)
  • L/S 36 tubing (Cole Parmer EW-96410-36 or EW-06434-36)
  • DIN adapter (Molded Products MPC855NS.375)
  • SNP-8 hose clamps (Cole Parmer EW-06832-08)
  • Nitrile or latex gloves
  • 0.5% Tween 80
  • 500 mL bottle, sterile
  • 1 Liter bottle, sterile
  • Sterile distilled water
  • Pliers
  • Sharpie
  • Appropriate Personal protective equipment (PPE)


Safety warnings
Sterile gloves should be worn during all steps of this procedure to prevent possible exposure to waterborne pathogens and prevent contamination of the ultrafilter. Since the system is under pressure, a face shield should be worn to protect eyes and mouth.
Before start
Make a fresh backflush solution:
  • Add Amount2.5 µL of Tween 80 to Amount500 mL sterile distilled water. Swirl slowly to dissolve the 0.5% Tween 80.

Prepare backflushing solution
Prepare backflushing solution
A Amount500 mL volume is required per ultrafilter to be backflushed. Make a fresh backflush solution as follows:

  • Add Amount2.5 µL of Tween 80 to Amount500 mL sterile distilled water. Swirl slowly to dissolve the 0.5% Tween 80.

Be patient, the Tween 80 can take a while to dissolve. Attempt to limit the amount of foam produced, but some foam is inevitable.
Backflush Station Set Up
Backflush Station Set Up
Assemble the filtration system by coordinating the position of the peristaltic pump, ultrafilter and backflush solution.

Tubing will flow from backflushing solution, through peristaltic pump, into the "top" of the ultrafilter. Backflush will then flow from the ultrafilter to a sterile beaker or bottle.
Cut enough fresh tubing (L/S 36) to run from the Amount500 mL vessel containing backflush solution, through the pump, into the side of the "top" of the ultrafilter.

Open the side port of the "top" of the ultrafilter by removing the cap. Push the freshly cut tubing on the side port all the way to where the port meets the ultrafilter. Feed the tubing through the peristatic pump and place the end of the tubing into the backflush solution.
Secure all of the connection points with zip ties or clamps as needed.
Cut another piece of fresh tubing (L/S 36) to run from the colored "bottom" part of the ultrafilter into a sterile vessel. Attach a DIN adapter and screw into the bottom of the ultrafilter which will feed into the sterile vessel.
Secure all of the connection points with zip ties or clamps as needed. Attach the tubing and feed the tube into the sterile vessel for collection of backflush solution that has flowed through the ultrafilter.

Ensure the 2 remaining caps on the ultrafilter are secured tightly so as to prevent leaks.
Once the tubing, pump, ultrafiltration cartridge, and sterile vessel have been set up, plug the appropriate power cord into the outlet in the back of the pump and the other end of the power cord into a battery source or outlet within the laboratory.
Backflush the filter
Backflush the filter
Determine the desired direction of water flow and toggle the switch for flow direction. Turn the pump on in the direction that pulls backflush solution towards the ultrafilter and then into the sterile vessel.
Continue pumping until no backflush solution remains in the Amount500 mL bottle and the flow out of the ultrafilter has slowed to a trickle.

It will be necessary to briefly pump air after the backflush solution is used up in order to get all the liquid out of the filter.
Turn off the pump and replace the lid on the backflush solution that has passed through the ultrafilter.
This is your sample. Place at Temperature4 °C until further analysis. Ensure this vessel is properly labeled.

Disassemble backflush station and ensure all used materials are placed in the appropriate biohazard waste containers.
The ultrafilter may be stored at Temperature4 °C if needed. Ensure all open ports (side and end) are capped to prevent any possible leaks)