May 01, 2023

Public workspaceAureococcus anophagefferens Virus (AaV)/Kratosvirus quantuckense viral particle count by SYBR Green staining and flow cytometry (CytoFLEX S Flow Cytometer Beckman Coulter) V.2

  • 1The University of Tennessee, Knoxville;
  • 2Rice University
Icon indicating open access to content
QR code linking to this content
Protocol CitationEmily E. Chase, Alex Truchon, Samantha R Coy, Steven W Wilhelm 2023. Aureococcus anophagefferens Virus (AaV)/Kratosvirus quantuckense viral particle count by SYBR Green staining and flow cytometry (CytoFLEX S Flow Cytometer Beckman Coulter). protocols.io https://dx.doi.org/10.17504/protocols.io.dm6gpj331gzp/v2Version created by Emily E. Chase
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: April 29, 2023
Last Modified: May 01, 2023
Protocol Integer ID: 81186
Keywords: cytoflex, flow cytometry, violet side scatter, Aureococcus anophagefferens, AaV, Floreovirus quantuckensis, viral particles, counting virus
Abstract
A method for obtaining viral particle counts of Aureococcus anophagefferens virus (AaV)/Floreovirus quantuckensis on a Beckman Coulter CytoFLEX S Flow Cytometry System (CytExpert software) using SYBR gold staining. Adapted from a protocol originally produced by Dr. Samantha Coy for PBCV-1, which was derived from Brussaard 2004; and *Brussaard et al. 2010.


Brussaard CP. Optimization of procedures for counting viruses by flow cytometry. Appl Environ Microbiol. 2004 Mar;70(3):1506-13. doi: 10.1128/AEM.70.3.1506-1513.2004. PMID: 15006772; PMCID: PMC368280.

Brussaard, C. P. D., J. P. Payet, C. Winter, and M. G. Weinbauer. 2010. Chapter 11: Quantification of aquatic viruses by flow cytometry. Manual of Aquatic Viral Ecology. Waco, TX:American Society of Limnology and Oceanography. doi:10.4319/mave.2010.978-0-9845591-0-7

*includes a published protocols.io: dx.doi.org/10.17504/protocols.io.dpj5km

*Virus name updated 29-APR-2023.
Guidelines
To achieve the intended results of this protocol a configuration using a violet laser must be setup for a CytoFLEX S Flow Cytometer. This protocol is based on a CytoFLEX with Violet SSC on the Violet laser at the 405/10 position. See the "Setting Up Violet Side Scatter (VSSC) Channel" section of the CytoFLEX manual.

It is important to note that viral particle counts alone (using this protocol) does not deduce the number of infectious particles.
Materials
Cytoflex S Flow Cytometer
CytExpert Software + desktop computer with appropriate specs
96 well CoStar [REF#3795] Assay Plates
1000mL pipette and tips (or alternatives based on sampling size volumes)
SYBR Gold Stain [REF#S11494]
Pre-prepared 40 μL 25% glutaraldehyde 2 mL tubes; one for each sample to be stained
Hot water bath


Appropriate PPE for fixing using glutaraldehyde (gloves, glutaraldehyde waste disposal, etc.)
Shealth fluid + cleaning fluid + Mili Q water (or alternative) [Cytometer maintenance only]



Safety warnings
All
Before start
Boot up the CytExpert Software and follow the start up protocol. Familiarity with the machine and its setting will permit necessary adjustments for your experiments and help ensure accurate results and interpretation.
CytoFLEX Experiment Setup
CytoFLEX Experiment Setup
Create a new experiment with sample acquisition settings as follows:
  • FSC 500
  • SSC 200
  • VioSSC 200
  • FITC 2000
  • PerCP 500
  • PB450 40 [placeholder; this could be set for a different dye than Pacific Blue shown here, this will not change the results]
  • Threshold: (manual) Violet SSC 3000

Set the flow rate to slow (10 μL/minute), the sampling to 60 seconds (not by events), and the display to 100 000 events. Under Advanced -> Event Rate Setting select "high" (this is not necessary if events per second are below 10,000).

Flow rate may need to be adjusted for a high abort percentage (>10%), both adjusting the rate (custom or otherwise) and diluting dense (e.g., a higher dilution series) samples will solve this problem. Users should also aim for a certain number of events per second, typically between 300–2000.

Settings can be adjusted in real time by selecting "run" on samples and changing the acquisition settings. The full CytoFLEX configuration of which this protocol is based off of is pictured below (Figure 1) for full repeatability.

Figure 1. Detector configuration setup for the CytoFLEX Flow Cytometer used to establish this protocol.
Customised measurement setups can be produced with histograms, density plots, etc. However, based on the above acquisition settings and a plotting of FITC-A and VioletSSC (both at log distribution) will accomplish detection of Floreovirus quantuckensis. See Figure 2 for this population. These results are achieved with the sampling and staining procedure detailed below.

Figure 2. CytoFLEX S Flow Cytometer discovery of Floreovirus quantuckensis viral particles. Events enclosed (i.e., gated) within the purple circle represent a population attributed to viral particles.

Sample Fixing and Staining (adapted from sources as described)
Sample Fixing and Staining (adapted from sources as described)
To fix samples: prepare 2.0 mL cryovials (this permits the option of long term storage by flash freezing in liquid nitrogen) for a 1.0 mL sample with a final concentration of 1% glutaraldehyde (EM Grade). Samples can include filtered lysate, or unfiltered lysate – which may include host cells (Aureococcus anophagefferens) and bacteria. Samples should incubate for 45+ minutes at 4 °C in the dark.

To facilitate easy sampling these can be done ahead of time. For a 1.0 mL sample 40 μL of a working stock of 25% glutaraldehyde can be added to each tube, followed by 960 μL of the sample to be fixed at sampling.
Dilute SYBR Gold [Invitrogen REF S11494; 10 000×] to 100×. It is recommend to maintain the working stock in the dark and on ice (reduces dye fading).

Aliquots can be stored at 4 °C for one month.
A serial dilution of the viral particle containing samples (sans stain) should be performed using 100 μL of fixed sample transferred over to 900 μL ASP12A media (or otherwise the same media used for your own cultures). At each step mix by pipetting up and down.

Depending on a priori knowledge you may want fully process (i.e. stain and observe) dilutions at 10-1, 10-2, 10-3, and/or 10-4. Likely, a subset only of these will be explored, however observing multiple dilutions should provide more confidence in viral particle counts.


Protocol
ASP12A Recipe for culturing Aureococus anophagefferens
NAME
ASP12A Recipe for culturing Aureococus anophagefferens
CREATED BY
Eric Gann


Serially diluted samples can now be stained by the addition of the working stain solution to a 0.5× concentration. Samples should be incubated at 80 °C for ten minutes in the dark.

For a 1.0 mL sample 5 μL of 100× SYBR Gold can be added to reach this concentration.
Serially diluted samples can be cooled for 5 minutes in the dark at ambient temperature (~19–21 °C). Upon completion of this incubation samples can be processed using the CytoFLEX.
CytoFLEX Sampling Process
CytoFLEX Sampling Process
250 μL of each fixed and stained viral particle containing sample to be measured are pipetted into a 96 well CoStar [REF#3795] Assay Plate round bottom plate (can be substituted) and loaded into the CytoFLEX.

Please see the CytoFLEX Manual for detailed instructions of each step and machine use. Sample volumes can be lowered if necessary after taking flow rates into account. Given that glutaraldehyde has been used in the fixing and staining process, gloves need to be wore when handling the plate (e.g., when loading the plate into the CytoFLEX).
Labelled samples can then be run sequentially using the "Auto Record". Upon completion data can be either exported (most conveniently as a .csv) or reviewed using the "Statistics" window according to all events and events within user defined gated populations. Population counts are achieved by number of events within the viral particle containing gates (as suggested in Step 2). Back calculating your serial dilution series will permit the estimation of viral particles contained in your sample at the point of fixing.

Note that these are viral particle counts and not a determination of infectious viral particles.