Public workspaceAureococcus anophagefferens population count, and relative size (Violet SSC) by flow cytometry (CytoFLEX S Flow Cytometer Beckman Coulter)

  • 1The University of Tennessee, Knoxville
Icon indicating open access to content
QR code linking to this content
Protocol CitationEmily E. Chase, Alex Truchon, Steven W Wilhelm 2022. Aureococcus anophagefferens population count, and relative size (Violet SSC) by flow cytometry (CytoFLEX S Flow Cytometer Beckman Coulter). protocols.io https://dx.doi.org/10.17504/protocols.io.q26g7yby9gwz/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: September 22, 2022
Last Modified: December 01, 2022
Protocol Integer ID: 70407
Keywords: cytoflex, flow cytometry, violet side scatter, cell size, cell counts, Aureococcus anophagefferens
Abstract
A method for obtaining relative cell size, population density, etc., of the brown tide algae Aureococcous anophagefferens by a Violet Side Scatter configuration on a Beckman Coulter CytoFLEX S Flow Cytometry System (CytExpert software).

Guidelines
To achieve the intended results of this protocol a configuration using a violet laser must be setup for a CytoFLEX S Flow Cytometer. This protocol is based on a CytoFLEX with Violet SSC on the Violet laser at the 405/10 position. See the "Setting Up Violet Side Scatter (VSSC) Channel" section of the CytoFLEX manual. While visualization of A. anophagefferens can be achieved without a violet laser, it will not be possible to determine relative cell size.
Materials
Cytoflex S Flow Cytometer
CytExpert Software + desktop computer with appropriate specs
96 well CoStar [REF#3795] Assay Plates
1000mL pipette and tips (or alternatives based on sampling size volumes)


Shealth fluid + cleaning fluid + Mili Q water (or alternative) [Cytometer maintenance only]



Before start
Boot up the CytExpert Software and follow the start up protocol. Familiarity with the machine and its setting will permit necessary adjustments for your experiments and help ensure accurate results.
CytoFLEX Experiment Setup
CytoFLEX Experiment Setup
Create a new experiment with sample acquisition settings as follows:
  • FSC 200
  • SSC 40
  • VioSSC 75
  • FITC 200
  • PerCP 150
  • PB450 1 [placeholder; this could be set for a different dye than Pacific Blue shown here, this will not change the results]
  • Threshold: (manual) PerCP 5000

Set the flow rate to medium (30 μL/minute), the sampling to 60 seconds (not by events), and the display to 100,000 events.


Note
Flow rate may need to be adjusted for a high abort percentage (>10%), both adjusting the rate (custom or otherwise) and diluting dense culture samples will solve this problem. Users should also aim for a certain number of events per second, typically between 300–2000.

Settings can be adjusted in real time by selecting "run" on samples and changing the acquisition settings. The full CytoFLEX configuration of which this protocol is based off of is pictured below (Figure 1) for full repeatability.


Figure 1. Detector configuration setup for the CytoFLEX Flow Cytometer used to establish this protocol.

Optional Create a combination of density plots and histograms best suited for your analyses. An example set up is provided in the image below (Figure 2), where A. anophagefferens populations are clearly achieved. It is recommended to set up a histogram with time on the x-axis (final histogram) to account for the machine's measuring consistency as sampling will often start out slower and then stabilize. This can be applied to any gated population you have selected by clicking the text at the top of the histogram and selecting the population within the dropdown menu.

Figure 2. CytoFLEX S Flow Cytometer discovery of Aureococcus anophagefferens cell populations. Lassoed (i.e., gated) populations represent A. anophagefferens classified events.

Sampling Process
Sampling Process
250 μL of each sample to be measured (e.g., A. anophagefferens culture) are pipetted into a 96 well CoStar [REF#3795] Assay Plate round bottom plate (can be substituted) and loaded into the CytoFLEX. After opening the "Plate" window and clicking "Add Plate", samples can then be labelled.

Note
Volumes can be reduced after taking into account the flow rate and sample timing.

Labelled samples can then be run sequentially using the "Auto Record". Upon completion data can be either exported (most conveniently as a .csv) or reviewed using the "Statistics" window according to all events and events within user defined gated populations. Population counts are achieved by number of events within the A. anophagefferens gates, and subsequent relative size of events (cells) can be achieved through Violet Side Scatter channel results.


Note
Relative size by Violet SSC cannot be achieved for every cell type (normally "larger" microalgae cannot), a priori comparisons with other cell size calculation methods (e.g., FlowCam) must be conducted.