It is a usual procedure calculated the enzymatic activity results using a theoretical factor achieved in reaction optimum conditions, described below:
Cuvette at 37 ± 0.2 0C, 10 mm light path.
If one of the correlated parameters is modified, it is recommended to apply an enzymes calibrator indicated by the reagent manufacturer. Labtest Diagnostica recommends Calibra series to perform the AST/GOT system calibration.
DA/minute(test or calibrator) = (A1-A2)/2
Factor: (Calibrator activity)/(DA/minute (test) x factor
AST activity (U/L) = DA/minute (test) x Factor
If all the correlated parameters are fulfilled, the theoretical factor (1746) can be applied.