Protocol Citation: Rachel Rachel A Howard-Till, Claudia E Osorio, Marcos Olivos, Bradley J Till 2020. A do-it-yourself low-cost agarose gel documentation and DNA quantification system . protocols.io https://dx.doi.org/10.17504/protocols.io.bdtpi6mn
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: March 17, 2020
Last Modified: June 23, 2020
Protocol Integer ID: 34383
Keywords: DNA quantification, Gel documentation, ImageJ, Blue light transilluminator, DIY molecular biology
Abstract
This protocol describes how to build your own agarose gel documentation system to create digital images for quantification of DNA stained with dyes that fluoresce under blue LED light. Free software and inexpensive materials are used that can save tens of thousands of dollars when compared to similar commercial products.
Guidelines
This protocol includes instructions for a self-made blue light box. Of course, commercially available blue light transilluminators may also be used, if available.
The materials listed for building the light box are meant to serve as examples only. We encourage the user to be creative in sourcing materials. Many of these items are available at local hardware or electronics stores. You may also upcycle or repurpose free materials (such as a cardboard box) if available.
SYBR™ Safe DNA Gel StainThermo Fisher ScientificCatalog #S33102
STEP MATERIALS
Prepare lambda DNA dilutions of 2, 4, 8, 16, and 32 μg/ml.
This protocol includes instructions for a self-made blue light box. Of course, commercially available blue light transilluminators may also be used, if available.
The materials listed for building the light box are meant to serve as examples only. We encourage the user to be creative in sourcing materials. Many of these items are available at local hardware or electronics stores. In addition to the items listed above, you may need some plastic cable ties, small screws, a drill, and soldering supplies (soldering iron of at least 30 Watts and soldering wire). If using aluminum strips to mount the LEDs, a small hacksaw is needed to cut them to the desired lenghts. A recycled cardboard box can be used as the light box housing.
We use a Nikon D3400 DSLR camera for image aquisition. Most digital SLR cameras are capable of taking images in raw format and can be controlled by the Entangle software. A camera producing Raw format images is recommended for digital quantification of gel images. Check the information on the Entangle website to find a list of suitable cameras https://entangle-photo.org/. If using an alternative camera, check the diameter of the lens and be sure to purchase the correct diameter of orange lens filter (Tiffen 21).
We recommend a laptop computer with a minimum of 4 GB RAM, running the free Linux Ubuntu operating system. We encourage reuse of old or otherwise unused computers. Outdated Windows computers can easily be wiped and have Ubuntu installed.
Price list for Gel Doc components as of 6/4/2020
Tiffen filter $46.26
Orange safety glasses, 1 pair $1.58
Nikon D3400 digital camera, renewed $372.00
Thermal grease paste, 30g $7.99
Electrical wire $10.99
Polycarbonate sheet $9.99
Aluminum lighting strip $23.99
3W High Power SMD LED COB Chip Lights
royal blue 440-450 nm with PCB, 10 pieces $6.75
3.9 Ohm 5 Watt 5% Wire Wound Cermet
Sandblock Resistor, 2 pieces $3.30
10K Ohm 5 Watt 5% Ceramic Resistor
5W, 2 pieces $3.30
12V 2A power source, 1 unit $4.17
Cardboard box, recycled $0
Laptop with minimum 4GB RAM, ~$300
Total $790.33
Protocol materials
Tiffen 55 mM 21 filter (orange)Amazon
Materials, Step 4.1
SYBR™ Safe DNA Gel StainThermo Fisher ScientificCatalog #S33102
Always follow laboratory safety procedures. Wear appropriate PPE, including lab coats, gloves, and safety glasses. While blue LED light is less dangerous than UV, the high intensity blue light can still be harmful to eyes, and it is recommended to wear blue-blocking safety goggles (see materials list).
Before start
Download and install all software according to instructions.
Prepare lambda DNA dilutions of 2, 4, 8, 16, and 32 μg/ml.
Pour a 1% agarose gel with enough lanes to run your samples + 5 lanes for lamda standards.
Build an LED transilluminator
Build an LED transilluminator
Build the LED transilluminator.
Cut a rectangular hole in a cardboard box to the desired size of your illuminated window. Make sure this hole is smaller than the clear plastic sheet that you will cover it with. In our prototype, we cut the hole to 15 x 30 cm.
Cut aluminum strips to serve as a base for the LED lights. These serve to dissipate heat from the LEDs. Drill holes in the strips to attach the LEDs. Make two strips with 3 LEDs for each long side of the box, and two strips with 2 LEDs for each short side.
Connect the LEDs, resistors, and power source according to the circuit diagram shown below. Solder the wires to the LEDs and resistors as shown in the photos below, then apply thermal grease to the back of the LEDs and fix them into place on the aluminum strips with small screws. Finally, poke holes through the cardboard box and use plastic cable ties to secure the aluminum strips holding the LEDs.
Close the box and place the clear plastic (polycarbonate or equivalent) sheet over the top of the box to cover the hole.
Install software.
Install software.
Prepare a computer for image aquisition and image analysis.
B. Install sofware to control the digital camera. This protocol uses Entangle for Linux that is available from the GNOME/Ubuntu Software repository.
C. Install spreadsheet software. If you have installed Ubuntu, the free LibreOffice Suite comes preinstalled. LibreOffice is avaiable for most opperating systems. If you are using a commercial software for text, presentations and spreadsheets, why not try a free alternative?
D. Install software to convert RAW images to TIFF format. Ufraw-batch is used in Linux. To install, open a terminal window (found in applications in Ubuntu) and type the command below. You will be prompted to enter your password.
DNA electrophoresis
DNA electrophoresis
1h
1h
Load 5 μl of your genomic DNA sample on a 1% agarose gel. Also run 5 μl aliquots of lambda DNA standards of 2, 4, 8, 16, and 32 μg/ml (ng/μl). This will allow you to create a standard curve to compare your sample to. Run gel approximately 20 minutes at 100 V. The DNA should be migrated into the gel 0.5-1.0 centimeter. Longer running times can spread the DNA band more, and reduce accuracy of quantification.
30m
Stain gel 30 minutes in 1X SYBR dye (Dilute concentrate 1 in 10,000 in gel running buffer, i.e. TAE or 0.5X TBE). Stain can be included in the gel matrix when pouring the gel, but it generally increases the background fluorescence. Staining solution is stable for at least two weeks and can be reused for several gels.
30m
Photographing gel
Photographing gel
Photograph the gel using a suitable camera that can produce uncompressed raw format or TIFF (.tif) images. Directions are given below for using a self made system and free software. Ideally, the gel documentation system should be set up in a dark room, or other small room or closet that can be made relatively dark. Alternatively, the camera and transilluminator can be covered with a thick cloth that blocks out light.
5m
Make sure the computer and camera are switched on. The orange filter
Tiffen 55 mM 21 filter (orange)Amazon
should be screwed to the camera lens. Make sure the camera is set to save raw images. Connect the camera to the computer with a mini-USB cable or whatever is appropriate for your camera. A camera icon should appear on the desktop to show that the computer recognizes the camera (when using Ubuntu). The camera can be held over the light box with a ring stand or tripod to allow hands free imaging.
Open the Entangle program. A window will pop up that says "Camera is in use". Click Yes to unmount the camera.
Activate continuous capture mode by clicking on the small button in the toolbar that looks like two rectangles stacked together. Zoom and focus the gel by manually adjusting the camera with the light box turned off. To view the gel: make sure you are wearing orange safety glasses, turn off any overhead lighting, then turn on the light box.
Under the camera settings, adjust the exposure time. Usually 1/16 to 1/8 sec exposure is sufficient. Capture image by clicking the camera button. Image will display in the window. Take additional exposures as necessary. Images often look darker after converting from raw format to .tif, so it is better to take a slightly brighter exposure than you think is necessary.
Convert images to TIFF format
Convert images to TIFF format
Convert image from RAW (e.g. NEF) format to TIFF, if necessary.
Move images to be converted into new folder. Open a Terminal window, cd into the folder containing the image files, and enter the following command line:
Measure band intensities
Measure band intensities
Measure band intensities.
Make sure Image J or FIJI is installed and running. Open the image of your gel using the File > Open drop down, Control + O, or drag and drop the image onto the toolbar.
Using the rectangle tool (top left button), draw a small box around the brightest band in your gel. Then move the box to the first band in your gel (it is easiest to work left to right). You move the box by placing the cursor over the box, and it should turn to an arrow. Click and hold, then drag to the desired position. You can also use the arrow keys to move the box.
Measure the number of pixels in the band by selecting Measure from the Analyze menu, or Ctrl + M. A Results window should open displaying Area, Mean, Min, and Max values. Repeat the measurement process for each band on the gel. The size of the measurement box (Area) should be the same for all bands. If you accidentally resize the box while moving it, you need to start again.
After you have measured all the bands, save the values by copying and pasting into a spreadsheet, or select Save as from the File menu (Ctrl + S), and the values will be saved as comma separated values that can be imported into spreadsheet software.
Analyze values in Spreadsheet.
Analyze values in Spreadsheet.
Create a standard curve in a spreadsheet software using the "mean" values (pixel densities) measured from the bands of the Lambda DNA standards in your gel. Chart these values against the known concentrations of the Lambda samples.
Highlight the data to be charted.
Select Insert > Chart. Use the chart wizard to create an XY Scatter plot.
Fit a trend line to the charted data.
Click on the chart, then click on a point in the chart to select. The point should turn green to indicate it is selected.
Select Insert > Trend Line from the menu. A window will appear. Select 2nd degree polynomial function. Click Show equation to display the equation of the line on the chart.
Use the polynomial equation to calculate the concentrations of your gDNA samples by plugging the measured densities (mean values) into the equation. Unfortunately you can't copy/paste the equation from the chart, so you have to enter it manually into the formula bar. If the loading volume of your lambda samples and your gDNA samples are the same, you don't need to adjust the values to determine the concentration. If you loaded more or less, multiply or divide by the appropriate factor to calculate the concentrations of your samples.